GE DewPro MMY 245 Installation And Operation Manual Download Page 6

DewPro MMY245

 

Installation and Operation Manual

 

GE General Eastern                                                                                                                          

 

4

1.2 System 

Description 

1.2.1  Portable Trace Moisture Analyzer 

The DewPro

 MMY 245 Moisture Analyzer is designed primarily as a portable moisture instrument 

that can be used in a variety of gaseous applications. The portable unit can be used to quickly spot 
check the performance of dryers for compressed air, natural gas. Checking pure gas supplies, accurately 
testing SF6 gas in high power switchgear and transformers, gasified LNG, breathing air or oxygen and 
many other applications are easy with the MMY 245 portable analyzer. 
This is a well balanced, compact, easy to use versatile analyzer complete with all the units of measure 
and connectability required to keep track of the moisture critical processes in your facility. The rugged 
DewPro

 MMY 245 case has been selected to provide the maximum protection for the internal 

components in most any industrial environment where it is used. 
With the MMY 245 Portable Moisture Analyzer, measurement time is typically 5 minutes. The sensor 
is kept in a very dry desiccated environment with the dry-down taking place while the instrument is 
inactive or being transported to the next measurement point. For complete instrument specifications, 
see Section 6. 

1.2.1.1 Versatility 

The variety of applications in which the DewPro

 MMY 245 can be used is limited only by its range of 

operation for moisture and pressure and by the chemical nature of the gas sample. The unit is designed 
to be used for dry samples having a dew point temperature below freezing at the pressure of the sample 
introduced to the unit. Some typical applications are listed below: 
Desiccant dried air 

Dried Natural Gas 

Argon gas for welding 

Nitrogen gas supplies 

Breathing air 

SF

6

 gas for switchgear 

Helium gas supplies 

Oxygen gas supplies 

Hydrogen cooling loops 

Instrument air 

Glove box gas purifier loops 

Ethylene Production 

Heat treating atmosphere 
generators 

Gas Bottle filling operations 

… and many others. 

The processes listed above often require many different units of measure and the MMY 245 is able to 
provide readouts in most of the common units necessary for these applications. The unit of measure is 
simply selected from the operating keypad. An optional pressure sensor will automatically correct for 
changes in pressure that occur in a given stream or when spot-checking several sample points 
throughout a facility or process. 

1.2.1.2 Safety 

The DewPro

 MMY 245 has been designed to meet or exceed the FM requirements for Class I, 

Division 1 environments for portable instruments. 

1.2.1.3  User Friendly Operation 

The normal method of measurement is to connect the DewPro

 MMY 245 to the process sample tap via 

a short flexible high-pressure hose that employs a quick-connect fitting. The sample valve on the 
process piping is opened to start a sample flow to the unit. After allowing the sample flow to purge the 
hose for 15-30 seconds, the handle on the side of the analyzer is rotated up to the vertical position. This 
directs the sample across the sensor. The power button is pressed and the readout provides the moisture 
value in the desired unit of measure.  
The analyzer is programmed using the keypad and a matrix to select the moisture unit of measure, 
temperature unit of measure and other operating parameters. An optional pressure sensor can provide 
real time correction for concentration units of measure. In lieu of the optional sensor, a pressure 
constant can provide the proper value with known pressure conditions. 

Summary of Contents for DewPro MMY 245

Page 1: ...DewPro MMY245 Installation and Operation Manual GE General Eastern DewPro MMY 245 Portable Moisture Analyzer Installation and Operation Manual GE Measurement Sensing Technologies GE General Eastern ...

Page 2: ...rize the system before connecting or disconnecting the hoses or fittings to the analyzer or at the process connection Disclaimer General Eastern Instruments Inc reserves the right to change or modify our products in appearance or performance specifications at any time and without notice Therefore information in this document is subject to change without notice and does not represent a commitment o...

Page 3: ...___________________________________________________________7 Basic Operation ________________________________________________________________8 2 1 Operating Controls ______________________________________________________________8 2 2 Display Elements ________________________________________________________________8 2 3 Basic Sampling or Spot Checking _______________________________________________...

Page 4: ...________________________________________________ 23 4 3 4 Closed Loop Connection ____________________________________________________________ 24 4 4 Field Validator Setup____________________________________________________________25 4 4 1 Matrix Orientation _________________________________________________________________ 25 4 4 2 Single Point Validation Function Guide_____________________________...

Page 5: ...contact the factory immediately Product Structure 1 Certification R Standard Y Other 2 Pressure Compensation R Basic configuration P Pressure sensor built in for ppm compensation MMY245 Accessories Part No Description 63005033 Carrying case including shoulder strap 63005032 Shoulder strap 63004006 Cable assembly MMY 245 to DewPro MMY 30 RS485 cable 63004005 Connector cable assembly RS485 DewPro RS...

Page 6: ...low Desiccant dried air Dried Natural Gas Argon gas for welding Nitrogen gas supplies Breathing air SF6 gas for switchgear Helium gas supplies Oxygen gas supplies Hydrogen cooling loops Instrument air Glove box gas purifier loops Ethylene Production Heat treating atmosphere generators Gas Bottle filling operations and many others The processes listed above often require many different units of mea...

Page 7: ... alkaline batteries are included and are recommended for replacement The internal sample system provides for purging the hoses after they are connected to the sample tap and then directs the sample across the sensor 1 3 1 Sensor Theory The Planar Gold Aluminum Oxide sensor is a capacitance based technique for detecting the moisture content in gaseous samples This sensor is manufactured using semic...

Page 8: ...3 4 1 Internal Construction All the wetted surfaces of the MMY 245 DewPro are electro polished non hygroscopic metal components of either stainless steel or anodized aluminum Introducing clean samples will prevent these surfaces from becoming contaminated 1 3 4 2 Avoid Liquid Samples Avoid ANY liquid exposure of the sample lines and the instrument Liquids will damage the unit and void the Warranty...

Page 9: ... of time can move large amounts of moisture into the measurement cell and eventually cause the sensor to drift 1 3 4 9 Piping Recommendations Connection the MMY 245 to the process or sample tap in the proper manner will insure the most accurate measurements Connecting tubing should be made from materials that are non hygroscopic and capable of handling the sample pressure Keeping the length of con...

Page 10: ...ll turn the unit on or off The HOME key will display the matrix position V0 H0 or alternately V4 H0 Pressing the V and H keys advances the vertical or horizontal position within the matrix and the display will show the contents of that position Pressing the and keys increments or decrements the value in the matrix cell described in the right side of the display These matrix and operating functions...

Page 11: ...the connecting tubing for a few minutes WARNING If there is any evidence of aerosols or the presence of condensate in the exhaust DO NOT PROCEED The unit is designed for gaseous use only and any liquid droplets or aerosols will damage the desiccant chamber 3 Turn on the power by pressing the power switch 4 Rotate the handle to the measure position vertical 5 Observe the rate of change bar graph in...

Page 12: ...t introduce samples at greater than the rated pressure of 150 psig 10 barg 2 Do not introduce samples of gases corrosive to aluminum or gold into the unit This will damage the sensor and will void the warranty 3 Do not introduce samples of known high moisture or samples saturated with water into the unit The sensor can be damaged and will not be covered under warranty Observe the outlet port to se...

Page 13: ... 10 X 10 matrix contains all the operating parameters for properly displaying and calculating the correct value in the appropriate unit of measure The control functions for the outputs and interface connections are easily set by moving through this matrix and selecting the appropriate values for each cell The following section describes the features and usage of the various matrix locations as the...

Page 14: ...V 2 hold Temperature A D V5 Scale Output 2 Opt Input Temp C 0 V Temp C 5 V bar 0 V bar 5 V Min Input Display Value Max Input Display Value Minimum Input Voltage Maximum Input Voltage Minimum Output Voltage Maximum Output Voltage V6 V7 Channel 2 Setup Output 2 Enable 0 Disable Output 2 D A Cal 0 V Output 2 D A Cal 5 V V8 Special Double Pt Dewcomp 1 begin Double Pt State 100 OK Single Pt DewComp 1 b...

Page 15: ... Double Point State 100 ok V8 H3 Single Point State 100 ok V8 H5 Battery Capacity Remaining V9 H0 Display Primary Error Code V9 H1 Display Secondary Error Code V9 H2 Unit ID V9 H3 Software Version 2 5 3 Function Guide Using the Matrix refer to Figure 5 The matrix is organized in functional families by rows The following is a general description of these functional families indicating the row in wh...

Page 16: ...led system reset performed using V9 H9 or unit of measure selection is changed V0 H3 Displays the highest moisture value measured since the unit was reset using V0 H4 new batteries were installed system reset performed using V9 H9 or unit of measure selection is changed V0 H4 Resets the event memories of V0 H2 and V0 H3 by pressing the button V0 H7 Fault status for output 1 is selected with 0 sett...

Page 17: ...n error code is displayed enabling the user to identify any software errors when communicating to the factory technicians See the trouble shooting section for details V9 H1 The error preceding the current error displayed in V9 H0 is indicated enabling the user to identify any hardware errors when communicating to the factory technicians V9 H2 The instrument is identified by a series ID number that...

Page 18: ...he displayed number can be increased or decreased using the and button to adjust the voltmeter reading to 0 V Value is around 500 This is factory calibrated V7 H9 Calibrates the output 2 hardware at 5 V connecting a voltmeter to the output the displayed number can be increased or decreased using the and button to adjust the voltmeter reading to 5 V Value is around 2500 This is factory calibrated 2...

Page 19: ...ut resistance of at least 20 KΩ to provide 0 5 accuracy A higher input impedance is recommended for better signal accuracy These outputs are scaled and referenced to specific units of measure using V1 H0 through V2 H7 for output 1 and V5 H0 through V5 H4 for output 2 See the section 2 5 5 above titled 2 5 5 Special Output and Input Functions 3 Troubleshooting Field maintenance consists of keeping ...

Page 20: ... Software errors V9 H0 Error Description 0 No Error 1 Dew point 100C 2 Dew point 20C 4 Voltage output on CH1 scaled to 0V 8 Voltage output on CH1 scaled to 5V 16 Temperature out of range 64 Voltage output on CH2 scaled to 0V 128 Voltage output on CH2 scaled to 5V An error indicating that the dew point is both 100C and 20C indicates that the output on CH1 cannot be calculated because of an error re...

Page 21: ...n is to verify whether a permanently mounted transmitter is functioning properly and in calibration and if not then apply the correct data to the EEPROM to realign its reading Although factory recalibration of the MMY 245 analyzer is recommended every 6 12 months depending on use as a spot check instrument it should be recalibrated within 4 months prior to MMY 3 series transmitter field validation...

Page 22: ...dition is indicated on the display that the unit is allowed another ten minutes to settle before continuing the validation procedure The two procedures are a single point validation that simply changes the EEPROM data to move the intercept of the curve and the two point validation that changes the slope and intercept The single point procedure is very brief requiring the sample readings of both de...

Page 23: ...gular plastic tab see Figure 7 4 Plug the small connector into the black socket on the PC board Figure 7 enlarged view 5 Arrange the wires so they won t get pinched by the plastic tab and secure the tab in place with the captive screw 6 Reinstall the display for display versions and the enclosure cover The unit is now ready for connection to the MMY 245 The cable connection may remain installed in...

Page 24: ...lidation If the orifice is at the outlet as in a model MMY 30R1A1A the setting for the pressure compensation in the MMY 245 must be set for proper pressure compensation In this case the sample pressure at the transmitter should be measured and noted for proper configuration of the validator If the orifice is at the inlet as in M N MMY 30R1B1A the setting for the pressure compensation in the MMY 24...

Page 25: ... validation procedure can begin The validation process should be allowed to equilibrate for a minimum of 3 hours at vacuum with the operation handle in the vertical position for best results Reverse this procedure to disconnect the MMY 245 When using the MMY 245 as a validator with a MMY 31 or MMY 35 at either positive or negative pressures use the parallel connection technique that follows 4 3 3 ...

Page 26: ...ted for proper configuration of the validator If the orifice is at the inlet as in M N MMY 30R1B1A there is no need to alter the setting for the pressure compensation in the MMY 245 Once the sample connection has been made the unit should be allowed to purge for at least 10 minutes to provide a true sample to the validator After purging the system the selected calibration may proceed by setting th...

Page 27: ... at the sample dew point This new value represents a change in the intercept of the calibration curve only A two point calibration is performed in the same manner as in a single point validation with the additional step of monitoring a second moisture value generally that of the ambient air The MMY 245 values are assumed to be the field standard and are used to reprogram the EEPROM data in the tra...

Page 28: ... 30 Calculating and reprogramming new values into the slave transmitter 100 Successful compensation confirmed 10 Compensation cancelled early 20 Communication error after 5 retries 30 Unrecognized slave transmitter type 4 4 3 Two point Validation Function Guide Performing a two point validation requires the operator to monitor the MMY 245 display as the validation steps are completed After complet...

Page 29: ...de present Suspected errors in readings should be dealt with first by insuring all connections are leak free and the unit of measure is correctly set The pressure compensation mode should be verified that it is set correctly for the application After these items are verified to be correct and if the unit still produces questionable readings contact the factory field service department by calling 8...

Page 30: ...puts Analog Output Two adjustable and assignable non I S 0 5 VDC Output range 100 C to 20 C adjustable Digital Interface RS 485 with proprietary protocol for calibration and communication to MMY 3 series loop powered transmitters Accuracy 2 C dew point temperature Measurement Range 100 to 20 C dew point temperature Options Pressure sensor Hose kit with quick connect fittings Carrying Case with sho...

Page 31: ...DewPro MMY245 Installation and Operation Manual GE General Eastern 29 7 Appendix ...

Reviews: