5-16
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Troubleshooting:
Data Acquisition Tests
Data Acquisition Tests
ECG Functions
1. Connect the Multifunction Microsimulator (pn MARQII) and
appropriate patient cables, to the ECG connector of the monitor.
Turn the monitor and the patient simulator on.
2. Set the monitor to display leads I, II, III, and V simultaneously:
a. From the main menu, select
DISPLAY OPTIONS
.
b. Make sure the Display menu item shows Individual plus six
waveforms. If it shows Full, change it to Individual.
c. Select
WAVEFORMS ON/OFF
from the menu.
d. Set the displayed waveforms for the following ECG leads:
3. Set the patient simulator to output calibration (cal) pulses at 1.0 mV.
4. Measure the cal pulse (
) amplitude. These should be (+/–20%):
Lead I:0.5 mV
Lead II:1 mV
Lead III:0.5 mV
Lead V:–0.5 mV
5. It may be necessary to run a graph to accurately measure the cal
pulses. Perform these steps to graph all four waveforms.
a. From the main menu, select
GRAPH & ALARMS
.
b. Select
GRAPH CONTROL
from the menu.
c. Set the graphed waveforms for the following ECG leads:
d. Press the
GRAPH GO/STOP
front panel control on the monitor
to start and stop a manual graph.
ECG 1
WAVEFORM 2
WAVEFORM 3
WAVEFORM 4
LEAD II
LEAD V
LEAD I
LEAD III
ECG 1
WAVEFORM 2
WAVEFORM 3
WAVEFORM 4
LEAD II
LEAD V
LEAD I
LEAD III
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......