Revision C
250cx Series Maternal/Fetal Monitor
15-5
2036946-001
Maintenance: Cleaning
Sporicidin* Soap and water
2.
Rub soiled area until clean.
3.
Dry with a soft, dry cloth.
Maternal NIBP Cuffs and Hoses
General
The cuff must be thoroughly cleaned with the specified detergent before reuse. The
additional use of household bleach as described below provides at least
intermediate-level disinfection.
Apply cuff hose plugs before cleaning.
The following cleansing procedure was repeated 20 times on DURA-CUF
®
Blood Pressure Cuffs and once on SOFT-CUF
®
Blood Pressure Cuffs
without affecting the performance of the cuff.
While this procedure is adequate for cleaning/disinfection, it may not
remove all stains.
Do
not
immerse hoses.
Do
not
immerse cuffs without prior application of cuff hose caps.
Materials
Enzymatic detergent such as ENZOL* enzymatic detergent (US) or
Cidezyme* enzymatic detergent (UK)
Distilled water
10% solution of household bleach (5.25% sodium hypochlorite) in distilled
water
Soft cloths and soft-bristled brushes
Spray bottles
Procedure
1.
Prepare the enzymatic detergent according to the manufacturer’s instructions
and the 10% bleach solution, in separate spray bottles.
2.
Spray the detergent liberally on device. If the material is dried on, allow the cuff
to sit for 1 minute. For soil on the soft part of the closure or the cuff itself, wipe
the material off with a soft cloth. For persistent contamination on the soft part of
the closure, use a soft-bristled brush to loosen particles. Rinse with copious
amounts of distilled water. Repeat until no visible contamination remains. For
soil on the hook part of the closure, use a soft-bristled brush to remove the
material, and rinse with copious amounts of distilled water. Repeat until no
visible contamination remains.
3.
Spray the 10% bleach solution on the affected area until the area is saturated.
Allow the cuff to sit for 5 minutes.
4.
Wipe away any excess solution and rinse the cuff again with distilled water.
Allow 2 hours for drying.
*
Trademarked
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......