5011A EN 20020315
CASSETTE READ/
WRITE ERROR
Press , remove
and try again
• Press Confirm key.
• Remove cassette.
• Re-insert cassette.
• If problem persists, initialize
and identify cassette via ID
Station.
• If problem persists with other
cassettes, contact your local
service organization.
CASSETTE NOT
IDENTIFIED
Press , remove
and identify
• Press Confirm key.
• Remove cassette.
• Identify cassette.
• Re-insert cassette.
24 X 30 CM
CALIBRATION
MISSING
Press
to accept
or
• Press Confirm key to treat
24 x 30 cm image plate
without calibration or press
Cancel key to treat cassettes
with other formats.
• Contact your local service
organization.
SERVICE MODE
Please wait
Wait.
CASSETTE SLOT
BLOCKED
Remove cassette,
press
• Remove cassette.
• Remove obstructing objects.
• Press Confirm key.
IMAGE-QUEUE FULL
Check queue
image transmission queue
(‘Queue management’)’
on
page
manual.
• Check that the Centricity™
SP1001 is not off line (Refer
to
on page
Status field:
LOCKED
MESSAGE 1
Message 2
Action