28
5011A EN 20020315
Reading an image plate in the dedicated configuration
In the dedicated configuration, the Centricity™ SP1001 can digitize the image
plate while you are entering the identification data via the ID Station.
To read an image plate:
1
Check that the Centricity™ SP1001 is ready for operation:
• the Centricity™ SP1001 must display the operator main screen with ‘Ready’
status, e.g.:
• the status indicator at the top of the Centricity™ SP1001 must be green and be lit
constantly.
❖
The Centricity™ SP1001 is operational if the status field equals ‘READY’,
even if status messages of the destination are shown (e.g. ‘QS not ready’).
2
Press the appropriate emergency button at the front of
the
Centricity™ SP1001.
The emergency button determines the speed class, i.e. the sensitivity, which will be
used to digitize the image plate.
The button which you have pressed will be lit and the display will read:
For digitizing images of the trunk.
For digitizing images of the limbs.
QS not ready
READY
SP1001
Status field
BUSY
DIRECT ID
IDENTIFY AT ID STATION
or press
to use emergency data
SP1001
5011 A UM EN CR SP1001.fm Page 28 Friday, March 15, 2002 4:11 PM