ECG
ECG module
Analog output signal
E-modules
ECG 1 Lead (top waveform position).
If ECG 1 Lead is any of the derived leads related to the 12RL 12 lead, then the
analog output will use lead II.
PDM
ECG 1 Lead (top waveform position).
TRAM
ECG 1 Lead (top trace position).
Tram-Rac
Lead II is the first ECG lead for analog output.
ECG 1 Lead is the second ECG lead for analog output.
Va is the third ECG lead for analog output.
About the combination monitoring mode
The combination monitoring is a licensed feature. In combination monitoring mode
ECG is acquired from a telemetry receiver system. This ECG data acquisition capability
enhances basic telemetry monitoring by providing access to all of the available
parameters from bedside monitors, while acquiring the ECG data from telemetry. In
this monitoring mode, all data — local and telemetry — is viewed at the central station
and the bedside monitor. However, any historical data stored at the central station will
be unavailable. Any new alarm history samples created on the telemetry transmitter
cannot be viewed on the monitor if they are created after the combination monitoring
has been started. Only the snapshots created on the monitor and the samples of
the telemetry transmitter created prior to starting the combination monitoring can
be viewed.
The combination monitoring mode can be enabled through
Monitor Setup
>
Default
Setup
>
Care Unit Settings
>
Telemetry
>
Monitor or Telemetry
, and the setting is
password protected. The monitor needs to be configured to the CARESCAPE Network.
This option cannot be used in the NICU software package.
In case the telemetry patient has been admitted when the device is connected to the
monitor, the arrhythmia alarm priorities and limit alarm priorities (except in case the
monitor alarm is set to escalating) and the following ECG settings of the telemetry
will be used:
●
HR, ST, and PVC alarm limits
●
PVC alarm status (on/off)
●
Pacemaker detection
●
Lead analysis
●
Va lead position
●
Primary lead
●
ECG waveform size
●
Arrhythmia detection level
●
ST analysis status (on/off)
When combination monitoring is started with a non-admitted telemetry patient, these
same settings from the monitor will be sent to telemetry. Additionally, the
Telemetry
Waveforms
printing location is sent to telemetry.
If the telemetry alarm priority is such that it is not supported by the monitor, it will be
mapped to the next higher priority available.
2062971-001
CARESCAPE Modular Monitors
155
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...