8838 User Guide (BB1857-H)

360° Images and the Stitch Angle

If you set the 

Extent

 parameter for a 3D image to 360°, a stitch angle appears on the 

image. (The stitch angle parameter is only active when 

Extent

 is set to 360°.)

Stitch angle vs

stitch line

The stitch 

angle

 is a parameter that you set on the screen. The stitch 

line

 is the line 

that is in the 3D image, reflecting the setting of the stitch angle parameter. 

By default, the stitch line is located at the bottom of the cube, which is normally away 
from the area of interest; at the bottom of the cube (opposite to the default position). 
To make sure that the stitch line in the screen does not interfere with what you want 
to see, check the cube after the 3D imaging. If necessary, rotate the stitch line to a 
new position.

Adjusting the

stitch angle

To adjust the stitch angle – on Flex Focus 1202:

1

Select the appropriate pro package for your imaging session and click 

Stitch 

in the 3D General tab.
In the default center position  ,

 

the stitch angle is opposite the array’s default 

position; that is, towards the transducer button (see Figure 6).

Figure 6. The stitch angle’s default position.

2

Select the position you want by choosing the icon that indicates the desired 
position – away from the area of interest.

A stitch angle setting applies only for a particular imaging session.

When you start the imaging, the control panel is locked to avoid accidental 
interruption of the imaging process. A progress bar is displayed at the top of the 
image as a graphic image and as the number of seconds remaining to show how much 
time is left to complete the image.

To adjust the stitch angle – on Pro Focus 2202 UltraView:

1

Select the appropriate pro package for your imaging session and click 

Stitch 

in the left menu bar.
In the default center position  ,

 

the stitch angle is opposite the array’s default 

position; that is, towards the transducer button (see Figure 6).

2

Select the position you want by adjusting the slider bar that indicates the desired 
position – away from the area of interest.

A stitch angle setting applies only for a particular imaging session.

The stitch angle’s default center position 
is towards the transducer button

Summary of Contents for BK MEDICAL 8838 Series

Page 1: ...For Professional Users Only English BB1857 H April 2023 User Guide Type 8838 3DART Transducer...

Page 2: ...products and services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for...

Page 3: ...ing and Disinfection 7 Starting Imaging 7 Connecting the Transducer 7 Changing Frequency 8 Using a Transducer Cover 8 Using the Transducer Control Button 9 Changing Orientation 9 Endovaginal Anorectal...

Page 4: ...4...

Page 5: ...suitable for contrast imaging1 Patient Population The patient population is adults Figure 1 3DART Transducer Type 8838 Imaging Plane The transducer contains a linear array of ultrasound elements that...

Page 6: ...with the patient Turn off the system Unplug the system from the wall and make sure it cannot be used until it has been checked Do not try to repair the system yourself Contact your BK service represen...

Page 7: ...n and responsibility to provide the highest possible degree of infection control to patients co workers and themselves The instructions in this book are meant as a guide To avoid cross contamination f...

Page 8: ...oluble agent inside and outside the cover to create good acoustic contact To put on the cover 1 Apply gel to the tip of the transducer so that it covers the entire imaging surface most of the black pa...

Page 9: ...e the orientation of the image on the monitor refer to the applicable system user guide for instructions To change the angle of the 2D imaging plane see Rotating the 2D Imaging Plane on page 10 Endova...

Page 10: ...the position of the transducer array Note that the default position of the imaging plane is opposite the transducer button see Figure 2 Figure 4 Rotation symbol on the flex Focus 1202 Not available o...

Page 11: ...to its default position You should always check the position of the imaging plane to make sure that it is in the desired postition 3D Imaging NOTE See the advanced system user guide for more informat...

Page 12: ...the stitch angle is opposite the array s default position that is towards the transducer button see Figure 6 Figure 6 The stitch angle s default position 2 Select the position you want by choosing the...

Page 13: ...change this parameter on the system Untracked 3D acquisition available with some transducers is not an option for this transducer Select the Mover parameter on the screen There are two options IP8838...

Page 14: ...lowed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling See the applicable system user guide for further details WARNING D w1 For disp...

Page 15: ......

Page 16: ......

Reviews: