Appendix E. CE Mark Compliance
Appendix E. CE Mark Compliance
E.1 Introduction
For CE Mark compliance, the PanaFlow™ LC flow meter must be wired in accordance with the instructions in
this appendix.
IMPORTANT:
CE Mark compliance is required for all units intended for use in EU countries.
E.2 Wiring
The PanaFlow™ LC must be wired with the recommended cable, and all connections must be properly
shielded and grounded. Refer to
below for the specific requirements.
Note:
If the PanaFlow™ LC is wired as described in this appendix, the unit will comply with the EMC Directive.
Table 33: Wiring Requirements
Connection
Cable Type
Ground Termination
Transducer
Armored RG62 A/U
Grounded using a cable gland.
Input/Output Armored 22 AWG shielded (e.g. Baystate #78-1197) with
armored material added to outside of jacket
Grounded using a cable gland.
Power
Armored 14 AWG 3 conductor
Grounded using a cable gland.