Maintenance Tasks
© 2018 Baker Hughes, a GE company – All rights reserved.
DPS 5000 Series Instruction Manual-English | 15
4. Maintenance
Note:
Until installation, keep the unit in the original container with all the covers in
position. The container and covers prevent contamination and damage. When not in use,
keep the connections clean at all times, and put the covers on the open connections.
4.1 Maintenance Tasks
• Identify faults from the software.
• Clean the unit.
• Recalibrate the unit.
Note:
BHGE can provide a calibration service that is traceable to international
standards.
Return the unit to the supplier for these items:
• For all repairs.
• For changes or upgrades to the internal software.
4.2 Maintenance – From the Software
Note:
0x identifies a hexadecimal value. Object 0x1800 02 = Index 0x1800, Sub-index
0x02 (refer to Appendix A., “CANopen Object Dictionary,” on page 19).
4.2.1 The Status of the Unit
To monitor the current status of the unit, use these sources of error data:
• The Error Register (object 0x1001) (refer to Appendix A., “CANopen Object
If there is an error:
• Do the Network Initialization Process (the boot-up process) again.
• Examine the electrical connections. Do all the applicable tests and checks (refer to
the System Installation Manual).
• If necessary, install a new pressure transducer.
WARNING
High pressures and temperatures are dangerous and can
cause injury (Refer to pressure limits in the sales data sheet). Be careful
when working on components connected to lines that have high
pressures and heat. Use the applicable protection and obey all safety
precautions.