5-6
ApexPro™
2001989-200E
Alarms
After five minutes, the LED on the transmitter will no longer flash and alarms will be
reactivated.
Reactivating alarms at the transmitter
To reactivate the alarms before the five minute time period has elapsed, press both
transmitter buttons simultaneously again.
Alarm pause breakthrough
NOTE
For more information on the
Alarm Pause Breakthrough
feature in
Combo
mode, refer to
Combo and Rover Combo monitoring on page 6-19
NOTE
The
Alarm Pause Breakthrough
feature defaults to
Always On
. It can be set to
Always
Off
in the
Telemetry Unit Defaults
tab (in service mode only) sheet
BEFORE admitting a telemetry patient if you do not wish to have
Crisis
level
alarms break through alarm pauses.
This feature cannot be set on an individual patient basis. It is either on or off for
all telemetry patients admitted to the CIC Pro center. A status message on each
patient’s
Alarm Control
tab sheet indicates whether it is enabled (on) or disabled
(off).
The
Alarm Pause Breakthrough
feature allows any
Crisis
level alarm to break
through (interrupt) an alarm pause and sound at the CIC Pro center.
In other words, when this feature is turned on in the
Telemetry Unit Defaults
tab
sheet,
Crisis
level alarms will sound at the CIC Pro center, even if an alarm pause is in
effect.
The chart below illustrates the function of the
Alarm Pause Breakthrough
feature
during the various alarm states.
After a
Crisis
level alarm has broken through an alarm pause, the telemetry system
does
not
return to an alarm pause state. All alarms at any alarm level will sound at the
CIC Pro center.
If you wish to continue pausing alarms after an
Alarm Pause Breakthrough
occurs,
you must re-initiate the alarm pause:
Alarm Pause State
Alarm Pause Breakthrough Feature Enabled
Alarms on
No
Alarm Pause Breakthrough
; all alarms are on.
Alarms off
No
Alarm Pause Breakthrough
; all alarms are off.
Alarm off reason (X-ray,
shower, etc.)
Alarms are paused;
Crisis
level alarms will break through
the alarm off reason if the patient is in antenna range.
Alarms paused from the
transmitter
Alarms are paused;
Crisis
level alarms will break through
the alarm pause.
Summary of Contents for ApexPro
Page 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Page 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Page 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Page 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Page 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Page 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Page 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Page 86: ...5 18 ApexPro 2001989 200E Alarms ...
Page 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Page 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Page 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Page 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Page 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Page 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Page 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Page 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Page 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Page 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Page 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Page 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Page 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Page 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Page 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Page 217: ......