B-4
ApexPro™
2001989-351A
Electromagnetic compatibility
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should not be used closer to any part of
the equipment, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
EN 61000-4-6
3 Vrms
150 KHz to 80 MHz
0.3 V rms
c
d
= 11.6
d
= 11.6
80 MHz to 800 MHz
d
= 23.3
800 MHz to 2.5 GHz
Radiated RF
EN 61000-4-3
3 V/m
80 MHz to 2.5 GHz
0.3 V/m
c
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the
transmitter manufacturer, and
d
is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1
: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects, and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and FM
radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment
is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the equipment.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 0.3 V/m.
c
The low power design does not allow the use of high dynamic range operational amplifiers. To further reduce size, the ECG section utilizes
single supply topology, and unshielded Flex assemblies. The result of the above mentioned design choices results in a design that provides
acceptable performance in the clinical environment.
Summary of Contents for ApexPro T14
Page 9: ...2001989 351A ApexPro vii ...
Page 10: ...viii ApexPro 2001989 351A ...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction ...
Page 20: ...1 10 ApexPro 2001989 351A Introduction ...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview ...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration ...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration ...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance ...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance ...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting ...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts ...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module ...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout ...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications ...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility ...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility ...
Page 191: ...2001989 351A ApexPro 1 1 ...