1.
Turn on the RF monitor and allow it to operate for at least one hour for
temperature stabilization.
2.
Set the center frequency to the frequency of the transmitter. Refer to the ApexPro
Telemetry Frequency Chart Reference Manual.
a.
Press
FREQ
.
b.
Press the numbers corresponding to the frequency of the transmitter.
c.
Press
MHz
and press
Enter
.
3.
Set the span to 500KHz (50KHz/div).
a.
Press
Span
.
b.
Press
500
.
c.
Press
KHz and press Enter
.
4.
Set the reference amplitude to -10dBm at 10dB/div.
a.
Press
AMPT
.
b.
Press
10
.
c.
Press
GHz/-dBm
and press
Enter
.
5.
Connect the transmitter with lead wires to a patient simulator.
6.
Wrap the leadwires around the antenna of the RF monitor.
316A
7.
Verify that a peak signal of at least -20dBm (-10dBm for T14 transmitter) can be
obtained. It may be necessary to move the transmitter and leadwires around to
obtain this peak signal.
Summary of Contents for ApexPro T14
Page 9: ...2001989 351A ApexPro vii ...
Page 10: ...viii ApexPro 2001989 351A ...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction ...
Page 20: ...1 10 ApexPro 2001989 351A Introduction ...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview ...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration ...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration ...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance ...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance ...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting ...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts ...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module ...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout ...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications ...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility ...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility ...
Page 191: ...2001989 351A ApexPro 1 1 ...