Action
Step
Select an exposure option in the Exposure: drop down menu:
4
•
Auto
•
Semi-Auto
•
Manual - click the Exposure time button and set an exposure time and
then click Set.
More details are found in Set exposure time, on page 68.
•
For Advanced exposure see Fluorescence RGB with Advanced exposure,
on page 151.
If Auto or Manual exposure was chosen, select Start.
5
Result: A progress bar appears on the screen indicating the time remaining
of the imaging process and the selected exposure option.
Proceed to the final step of this instruction.
If Semi-Auto exposure was chosen, select Next and wait for the pre-exposure
to finish.
6
Result: A pre-exposed image appears.