![GE AKTAprocess Site Preparation Manual Download Page 6](http://html1.mh-extra.com/html/ge/aktaprocess/aktaprocess_site-preparation-manual_4080661006.webp)
Space requirements
The following space requirements must be fulfilled at the location where the ÄKTAprocess
system will be installed:
Specification
Parameter
2050 x 2250 mm
Minimum floor area for operating the
ÄKTAprocess 6 mm, 10 mm 3/8
″
and 1/2
″
system
(Cabinet 10)
(including 1 m of free space around the
system to allow service and mainte-
nance)
2050 x 3200 mm
Minimum floor area for operating the
ÄKTAprocess 1
″
SS and PP system
(Cabinet 30)
(including 1 m of free space around the
system to allow service and mainte-
nance)
Hori/- 1°
Floor inclination
WARNING
Access to power cord. Do not block access to the power cord. The
power cord must always be easy to disconnect.
6
ÄKTAprocess Site Preparation Guide 29-1396-55 AA
2 Site preparation overview