Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual
4-27
5610739-100 English Rev.10
B mode Checks
Introduction
The B Mode is the system’s default mode.
B Mode Screen Example
Preparations
•
Connect one of the probes.
•
Turn ON the scanner.
The B Mode is displayed (default mode).
Adjust the B mode controls
Press
B
on the Operator Panel to access B mode.
These Image Controls are used to optimize the B Mode picture.
Verify that all the listed controls are working as intended:
• Use
Gain
and TGC controls to optimize the overall image
together with the Power control.
• Use
Depth
to adjust the range to be imaged.
• Use
Focus
to center the focal point(s) around the region of
interest.
• Use
Frequency
(move to higher frequencies) or
Frame rate
(move to lower frame rate) to increase resolution in image.
• Use
Frequency
(move to lower frequency) to increase
penetration.
• Use
the
Reject
control to reduce noise in the image.
•
Use the DDP control to optimize imaging in the blood flow
regions and make a cleaner, less noisy image.
• Use
UD Clarity
(Cardiac) or
UD Sp. Red.
(non-cardiac) to
reduce image speckle. Extra care must be taken to select
the optimal Speckle reduction level, as too much filtering of
speckle can mask or obscure desired image detail.
• Use
Adaptive reject
(Cardiac) to reduce near field haze
and blood pol artifact without diluting tissue appearance of
moving structures.
Summary of Contents for 5661242
Page 5: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 3 5610739 100 English Rev 10 ...
Page 6: ...i 4 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 7: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 5 5610739 100 English Rev 10 ...
Page 8: ...i 6 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 9: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 7 5610739 100 English Rev 10 ...
Page 10: ...i 8 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 11: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 9 5610739 100 English Rev 10 ...
Page 12: ...i 10 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 13: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 11 5610739 100 English Rev 10 ...
Page 14: ...i 12 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 15: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 13 5610739 100 English Rev 10 ...
Page 20: ...i 18 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 28: ...i 26 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 58: ...Introduction 1 30 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 190: ...Service Adjustments 6 20 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 316: ...Index 4 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 ...
Page 317: ...GE ...