Section 10-4 - Tools Required
10 - 9
10-4-7
Grounding Continuity
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
The ground wire resistance should be less than 0.2 ohms. Reference the procedure in the IEC 601-1.1.
10-4-7-1
Meter Procedure
Follow these steps to test the ground wire resistance.
1.) Turn the LOGIQ C3/C5 Premium unit OFF.
2.) Plug the unit into the meter, and the meter into the tested AC wall outlet.
3.) Plug the black chassis cable into the meter’s “CHASSIS” connector and attach the black chassis
cable clamp to an exposed metal part of the LOGIQ C3/C5 Premium unit.
4.) Set the meter’s “FUNCTION” switch to the RESISTANCE position.
5.) Set the meter’s “POLARITY” switch to the OFF (center) position.
6.) Measure and record the ground wire resistance.
CAUTION
Electric Shock Hazard. The patient must not be contacted to the equipment during this test.
Figure 10-2 Ground Continuity Test
Summary of Contents for 5478035
Page 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10 ...
Page 2: ......
Page 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii ...
Page 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii ...
Page 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv ...
Page 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v ...
Page 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi ...
Page 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii ...
Page 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii ...
Page 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix ...
Page 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x ...
Page 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii ...