GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
55344303-100, R
EVISION
3
VIVID P3 S
ERVICE
M
ANUAL
7-20
Section 7-6 - Troubleshooting
7-6-1-9
Trackball
Figure 7-16 Trackball
The trackball is low sensitive!
Remove the ball from the Trackball,
and clean it.
YES
NO
Doesn’t the trackball move
at all?
END
The trackball module is defective
and replace it in Key board.