GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
2-6
Section 2-2 - General Console Requirements
2-2-4
Probes Environmental Requirements
Table 2-13 Operation and storage Temperatures for 2D Probes
Operational
Storage
Transport
Temperature
10 - 40
o
C
50 - 104
o
F
-10 - 60
o
C
14 - 140
o
F
-40 - 60
o
C
-40 - 140
o
F
Humidity
30 - 85%
non-condensing
30 - 90%
non-condensing
30 - 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Table 2-14 Operation and storage Temperatures for 4D Probes
Operational
Storage
Transport
Temperature
18 - 30
o
C
64- 86
o
F
-10 - 50
o
C
14 - 122
o
F
-10 - 50
o
C
14 - 122
o
F
Humidity
Max. 70%
non-condensing
Max. 90%
non-condensing
Max. 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa