Maintenance
3.
Remove the cover from the base.
See
“Replacing the Base Cover ” on page 28
for instructions.
4.
Remove the hex nut from the caster to be replaced.
Discard the caster. If the hex nut is undamaged, it can be retained to use on
the new caster.
5.
Repeat this procedure for the other casters.
6.
Reverse the disassembly procedures to reassembly the trolley.
7.
Reattach the MAC device and trolley options.
Replacing Casters on the MAC 1600 Trolley
This procedure applies only to the MAC 1600 Trolley. For instructions on removing the
casters from the Modular MAC Trolley, refer to
“Replacing Casters on the Modular
MAC Trolley” on page 28
.
Before removing the base, remove the MAC unit and accessories from the trolley. See
“Removing the MAC Unit from the MAC 1600 Trolley” on page 23
for instructions on
replacing the MAC unit to the MAC 1600 Trolley.
2044854-012E
MAC™ ECG Trolley
29