Functional and electrical safety checks:
Functional Checkout procedures
8-22
Dash 3000/4000/5000
2000966-542D
ECG tests
Equipment
Use the following equipment for these tests:
A multiparameter patient simulator.
ECG patient cable
ECG leadwire set
5 leadwire ECG test
Perform this test if the patient monitor acquires 5 leadwire ECG data.
Connections
1. Attach the ECG patient cable and ECG leadwire set to the
ECG/
RESP
connector on the patient monitor and connect the leadwires to
the patient simulator.
Patient Simulator Configuration
2. Set up the patient simulator as follows:
Heart rate – 80 bpm.
Heart rate amplitude – 1.0 mV.
5-leadwire ECG patient cable properly attached.
3. Admit the patient to the monitor.
Normal Sinus Rhythm Procedures
1. Observe the following:
ECG lead II is displayed and is noise-free
Heart rate of 80 ±1 bpm is displayed
With QRS tones enabled, an audible tone sounds with each
Rwave (QRS complex)
2. Verify ECG leads I, II, III, aVL, aVF, and aVR are available to view
and are noise-free.
Pacemaker Detection Procedures
1. Select
DETECT PACE
and set to
PACE 2
.
2. Select a pacemaker pulse on the simulator.
3. Observe the following while you view ECG leads I, II, III, aVL, aVF,
and aVR:
a “P” appears above the PVC count indicating pacemaker pulse
detection is enabled, and
the heart rate still reads 80 ±1 bpm.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......