Power Off
To power down the system, press down on the Power On Power Off/Stand-by switch.
After turning the power off, the system takes a few seconds to update certain parameters to the
hard drive before the power is disconnected.
NOTE: During this time the message
“Do not pull Power Cable.”
“Do not turn off Breaker.”
flashes on the monitor. The message “WARNING: NOW START THE POWER OFF
PROCESS.” is also displayed at the bottom of the screen.
While the system is running during this power down process, do NOT turn off the
circuit breaker in the back of the machine and do NOT unplug the system from the
wall outlet.
If the system has not turned off five minutes after pressing the power switch off,
listen for hard drive activity. If there is no hard disk drive activity, the circuit
breaker on the bottom of the power supply can be used to turn off the system. Do
NOT turn off the circuit breaker while the hard disk is working.