Revision A
Model 340 Telemetry System
5-13
2006920-001
Calibration: Transmitter Calibration
Main Oscillator
Connect the frequency counter to TP7. Adjust C135 for a frequency of
4.604 MHz ± 5 Hz.
Power Supply
1.
Confirm –5.6 V ± 0.2 V at TP11
2.
C5.8 V ± 0.2 V at TP10
3.
C5.0 V ± 0.25 V at TP16
4.
C2.50 V ± 0.05 V at TP12
5.
Confirm –2.50 V ± 0.05 V at TP14