System Setup
ECG Acquisition (cont'd.)
Field
Description
Prompt Level
Determines the level at which the
Hookup Advisor
prompts users
regarding patient hookup quality. This field is available only if
Hookup Advisor
is enabled.
There are three possible settings:
• Never
The system will not prompt the user when the signal quality
is poor. However, the statement
***Poor data quality,
interpretation may be adversely affected
will print on reports
for red lead quality situations.
• Yellow
The system will prompt the user for both red and yellow signal
quality situations. The user can then choose to continue or cancel
the ECG. The statement
***Poor data quality, interpretation
may be adversely affected
will print on reports.
• Red
The default value, the system will prompt the user for red signal
quality situations. The user can then choose to continue or cancel
the ECG. The statement
***Poor data quality, interpretation
may be adversely affected
will print on reports.
NOTE:
If
Hookup Advisor
and
Preview before analysis
are both on,
this setting is disregarded because any lead quality message
will be displayed when the user is prompted to continue in the
preview screen.
The generation of the statement
*** Poor data quality,
interpretation may be adversely affected
is based on the
Hookup Advisor
quality level as previously described even if
Hookup Advisor
is not turned on. If
Hookup Advisor
is not
turned on, the statement will be generated based on what the
Hookup Advisor
level would have been had it been enabled
Pre-acquisitio
n
Determines whether the system begins to acquire ECG data as soon
as the leadwires are connected to a patient. If set to
Yes
, the system
does not wait for the user to press
ecg
. The latest 10 seconds of ECG
data is ready for analysis when this setting is enabled.
ECG Analysis
To configure the analysis of ECGs, log on to the
System Setup
screen, select
ECG >
ECG Analysis
, and complete the fields described in the following table.
ECG Analysis
Field
Description
Preview before
analysis
Determines whether the system will display the ECG data before it
analyzes it. Applies for resting, pediatric, 15–lead, and vector loops
only. Select
Yes
to enable preview.
Screening criteria
Prevents specific 12SL statements from appearing on ECG reports.
Refer to the
12SL Physician's Guide
for more information on 12SL
statements.
2046275-016B
MAC
®
5500/MAC
®
5500 HD
125
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......