- 6 -
Figure 6 the subframe of user’s set
In Figure 6 Choise which one you need to change and move cursor “
√
” to there ,
then press “+” or “-” to Change.
1) Wait : Default 15min, range 1-15 min ( the increment Δ=1)
2) Pause: Default 5min, range 1-10 min ( the increment Δ=1)
3) Stir: Default 10s, range 5-90s ( the increment Δ=5s)
4) MTV: Default 80KV, range 10-100KV ( the increment Δ=10KV )
when voltage has been raised MTV (maximun test voltage) and keep 50 seconds with
no breakdown, the result of MTV will be the breakdown voltage of the electric insulating oil.
5) Breakdowns: Default 6, rang 1-6 ( the increment Δ=1)
After the choice press “OK” to start testing.
7. In (Figure 3) press”Opt” move cursor to “manual” press “OK” you will see need
password, you can contact with us to tell you.
press the "print" or "display" button, enter the oil sample single hit (Figure 2), press the
"OK" button, The storage record shows the subpage (Figure 7) by storing the voltage value
and the average value.
Figure 7 shows the subpage
8. The breakdown voltage and the number of cycles are automatically stored each time.