- 9 -
Check and adjust the gap between the electrodes make it back to standard values.
Observe surface of the electrodes with the magnifying glass. If derty, should scrub them
with cloths make them rehabilitate.
6. Instrument maintenance and debugging must be done by professionals
7. Before power on, you must check out the ground wire fastener fast and secure and
shell of instrument must be connected the ground wire!
8. After power on, Don’t touch the case cover of test cells to refrain dangerous shock.
9. Shut off power, During woking abnormal.
Ⅶ、
Fault code and removal
1. No work: Check the power wire and protective tube
2. No Voltage boot : Check the lip of test cell is or isn’t shut
3. No breakdown, voltage boot normally : Check the MTV opt(maximun test voltage)
4. No show after breakdown: Check test cell, clean the cell
5. No Print : Check if there is paper in printer.
6. Put on paper : we have put the paper in the printer. If paper is done , you must put new.
Operating instruction:
1
Press circle key on the printer front lip.
2
Put paper into printer. Keep enough paper out of the printer.
3
Close the lid.
Ⅷ、
Cup cleaning
1. Clean Cup
1
Use clean cloth wipes the surface of the electrode and the electrode rod;
2
Adjust electrode clearance with standard gauge;
3
Wash it 3~4 times with absolute alcohol and blow dry with a blower. Then use the
test oil to clean 2~3 times;
2. Clean stirring paddle
1
Use clean cloth wipes the surface of impeller. Don’t touch it by hand.
2
Use tweezers put the padddles in anhydrous alcohol for 2~3 times, and then blow
dry with blower