- 26 -
stop the alarm,and enter the next state.
4
Interface of remote control:This interface can be matched with the test gun and remote
control test stick.
5
Output voltage regulating knob:In the test process,adjust the knob clockwise,the output
voltage will increases;adjust the knob anticlockwise,the output voltage will decreases.Please
confirm this knob whether is in the 0 place when in the boot.
6
Current measuring return port:This terminal is the input port of the measuring current.In
the test process,we must pay attention to this terminal can’t fall off,if it fall off,the test object will
full of high pressure,that it may cause the occurrence of electric shock accident.
7
DC high voltage output port:The DC high pressure output port of test instrument will have
high voltage output when in the test,please don’t touch the high pressure port,the high voltage
output line, the high pressure probe alligator clip and the measured object.After the test is
finished,please confirm that the whole circuit is discharged completely then can replace the test
object.
8
AC high voltage output port:The AC high pressure output port of test instrument will have
high voltage output when in the test,please don’t touch the high pressure port,the high voltage
output line, the high pressure probe alligator clip and the measured object.
9
Power detection indicator lamp,"power supply grounding,safety"lamp:Connecting the
instrument and the power line use by city electric,the power switch of the tester is placed in the OFF
state;if the
10
connection of N,L,G in city electric is correct,the power ground safety indicator lamp will
light;if the connection of N,L,G is wrong,then the green lamp doesn’t light,please check the power
supply.
11
High voltage output indicator lamp:This lamp is out of control of the electric circuit.If the
voltage in the high voltage output port is greater than 110V,this lamp will flash.If in the reset
state,the lamp will flash,then the high voltage output port may have high output,it is prohibited to
touch any part of the test circuit.
12
Super leakage indicator lamp This lamp is on when the test is not qualified.
13
Test indicator lamp This lamp is on when the start button is pressed.
14
Voltage display window
15
Current display window:Time display window:The range of time is 0.0s~999s.When the