background image

The actual sensor value can be seen on the display rising or falling respectively. The magnets are the equivalent of 

the PCB UP and DOWN push buttons.
LOW ALARM is the next menu and left and right magnets increase and decrease this value. 
HIGH ALARM is next followed by OVER RANGE alarm. 
The direction of the alarms is displayed as ^ for rising and v for falling but these can be changed using left and 
right magnets together.

Appendix 1

Hyperterminal Display 

OHLDFI      

• 

O

 

H

 and 

L

 represent the Over Range,  High and Low alarms respectively ^^v   shows the direction of alarms

• 

D

 indicates if a duplicate address is detected 

• 

F

 indicates a fault 

• 

I

 indicates alarms inhibited 

Pressing

 (R)

 when in normal runtime display causes a sensor reset to occur. Gas type with address and serial 

number are then output to the PC together with alarms and calibration date, etc.  
This also  starts 1 minute autozero, which is useful when adjusting VR2 zero potentiometer. 
Pressing 

(C)

 on the PC will list the available commands

(V) 

Allows a view of the gas log taken at one minute intervals over 48 hours

(A) 

To set the sensor address

(G)

 Select gas type Select the gas type from a list

(Z)

  Prompts to use the zero potentiometer

(S) 

Prompts to use the PCB buttons

($)

  Initialise the sensor

(%) 

Clear the log of sensor readings, (this takes a minute to perform)

(U) 

Alarm directions, Rising or falling

(D) 

Enter calibration date

(R)

  Enter the range of the gas x 10

(F)

  Set 4mA level Use H and L and space

(T) 

Set 20mA level Use H and L and space

(O) (H) (L)

 Set over, high and low alarm levels x 10

(M)

 Set the value of full range raw processor measurement value when maximum gas was applied 

(E)

  Allows the editing of the 8 character user gas description

(X) 

Exit to normal display

Summary of Contents for XDI-XDIwin / 30J

Page 1: ...T EXCEED THE REQUIRED VOLTAGE OTHERWISE PERMANENT DAMAGE WILL OCCUR 6 Insert jumper G2 for a nominal gain value G1 is lowest gain G5 is highest or a combination G1 to G5 7 Connect a PC RS232 port using a micro USB lead or for earlier models a sensor programming module and a USB to RS232 converter part no 160510 and part no 160520 use hyperterminal set to 4800 1stop 8bit and initialise the sensor u...

Page 2: ...reed switches which can be activated using external magnets through the glass window of the flameproof XDIwin enclosure These magnets do not act instantly and have to be in close proximity to L M and R on the front display for a few seconds to activate a software setup function The right magnet allows the CAN address of the sensor to be changed When the address menu is displayed with a prompt to r...

Page 3: ...dress and serial number are then output to the PC together with alarms and calibration date etc This also starts 1 minute autozero which is useful when adjusting VR2 zero potentiometer Pressing C on the PC will list the available commands V Allows a view of the gas log taken at one minute intervals over 48 hours A To set the sensor address G Select gas type Select the gas type from a list Z Prompt...

Page 4: ...ID sensor can react to gas flow and humidity When gas is removed from the sensor the displayed PPM gas value may rise and take some time to return to normal Autozero is therefore delayed by 2 minutes after the reading has fallen below 5 of range ...

Page 5: ...re electrode stack if you are working in a dirty environment If the PID cell shows signs of contamination after the lamp window has been cleaned or is known to have been subjected to severe contamination it should be replaced It is recommended that the PID is recalibrated after the stack is replaced Instructions for replacing the electrode stack are below When do I replace the PID lamp A PID lamp ...

Page 6: ...the tips of cotton buds that are to be used to clean the lamps as this may contaminate them with finger oil 6 Ensure the lamp is completely dry and remove any visible signs of contamination before refitting Discarding the PID electrode stack The electrode stack does not have any toxic components however if it has been contaminated by toxic materials show due care when disposing Re fitting the PID ...

Reviews: