background image

T E C H N O L O G I E S

Technical Sheet ref C1518C v.2

292 D6C

XDI-XDIwin / 30J

PID SENSOR

  NEW SENSORS ARE SUPPLIED READY TO CONNECT AND GO  

BUT NEED 15 MINUTES POWER STABILISATION. The following procedure is for full set up 
using a PC but onsite calibration without PC/LCD requires only sections 18, 19, 20.

Note:

 If a new cell is fitted then full calibration is required. PID lamp has an approximate lifetime of 12 months. 

Replacement bulb and extractor tool can be purchased using part numbers 010-644 and 010-647 respectively.

PC setup

The auto zero adjust is continuous during power up inhibit period and PCB fault LED flashes to indicate this,  

do not apply any gas at this time because zero adjust will try to offset the measurement. Thereafter zero is 

adjusted every 60 seconds. D1 flickers orange at the zero point. This function can be toggled on/off by using 
command ‘Y’ in hyperterminal. ‘Auto zero’ is only active if gas reading is below 5% of full scale.

1.  Insert jumpers J20 and J6 position SO for 4-20mA source output
2.  Turn sensor voltage potentiometer RV4 fully anticlockwise (minimum voltage)
3.  Connect sensor to J2 terminal (W) white (P) pink (Y) yellow
4.  Connect 24v + and 0V to J10, insert test link at J9 to bypass 4-20mA terminal
5.  Set the sensor voltage to 5.0V measured across SV and 0V, adjust by slowly turning RV4 clockwise. 

DO NOT EXCEED THE REQUIRED VOLTAGE OTHERWISE PERMANENT DAMAGE WILL OCCUR 

6.  Insert jumper G2 for a nominal gain value (G1 is lowest gain, G5 is highest or a combination  

G1 to G5)

7.  Connect a PC RS232 port using a micro USB lead or for earlier models a sensor programming module 

and a USB to RS232 converter (part no. 160510 and (part no. 160520) use hyperterminal set to 
4800/1stop/8bit and initialise the sensor using (C) calibration mode, then shift + ($) command from the 
keyboard. The default gas is Isobutylene 200ppm.

8.  The typical hyperterminal display is (also see appendix 1)
 

60 C4H8 200ppm

 

O H L D F I

 

^ ^ ^

 

- - - - - -    PPM=0  A=010  R=010 Adj=500 Pinh=0 adj=4 CG=100 CGADC = 222

 PPM

  

is the gas value

 

A

 

is the processor raw measurement

 

R

  

is the zero point of the sensor (10 ± 2)

 Adj

 

is the adjustment value of autozero circuit, it can be between 0 and 1000 but 500 is ideal.

 

Pinh

 

is the number of seconds left of powerup inhibit (autozero is active during this time)

 adj 

 

is the number of seconds until a 1 second autozero will occur.

 

CG

 

is the calibration gas value PPM.

 

CGADC 

  is the number A must reach to display the value CG as the correct PPM gas reading

9.  Sensor Zero is true when D1 flashes orange and potentiometer VR2 is adjusted to ensure Adj (screen text) 

is between 400 and 600. Turning VR2 clockwise will increase Adj. Note: pressing (R) or J28 short will reset 
sensor and give 1 min adj time (when in clean air).

10.  Press F to adjust the 4mA output by measuring the mV across TP8 and TP9. Press H or L to adjust then press 

SPACE to set and exit. 

 

www.gds–technologies.co.uk

Summary of Contents for XDI-XDIwin / 30J

Page 1: ...T EXCEED THE REQUIRED VOLTAGE OTHERWISE PERMANENT DAMAGE WILL OCCUR 6 Insert jumper G2 for a nominal gain value G1 is lowest gain G5 is highest or a combination G1 to G5 7 Connect a PC RS232 port using a micro USB lead or for earlier models a sensor programming module and a USB to RS232 converter part no 160510 and part no 160520 use hyperterminal set to 4800 1stop 8bit and initialise the sensor u...

Page 2: ...reed switches which can be activated using external magnets through the glass window of the flameproof XDIwin enclosure These magnets do not act instantly and have to be in close proximity to L M and R on the front display for a few seconds to activate a software setup function The right magnet allows the CAN address of the sensor to be changed When the address menu is displayed with a prompt to r...

Page 3: ...dress and serial number are then output to the PC together with alarms and calibration date etc This also starts 1 minute autozero which is useful when adjusting VR2 zero potentiometer Pressing C on the PC will list the available commands V Allows a view of the gas log taken at one minute intervals over 48 hours A To set the sensor address G Select gas type Select the gas type from a list Z Prompt...

Page 4: ...ID sensor can react to gas flow and humidity When gas is removed from the sensor the displayed PPM gas value may rise and take some time to return to normal Autozero is therefore delayed by 2 minutes after the reading has fallen below 5 of range ...

Page 5: ...re electrode stack if you are working in a dirty environment If the PID cell shows signs of contamination after the lamp window has been cleaned or is known to have been subjected to severe contamination it should be replaced It is recommended that the PID is recalibrated after the stack is replaced Instructions for replacing the electrode stack are below When do I replace the PID lamp A PID lamp ...

Page 6: ...the tips of cotton buds that are to be used to clean the lamps as this may contaminate them with finger oil 6 Ensure the lamp is completely dry and remove any visible signs of contamination before refitting Discarding the PID electrode stack The electrode stack does not have any toxic components however if it has been contaminated by toxic materials show due care when disposing Re fitting the PID ...

Reviews: