GDS-68XP Operation & Maintenance Manual, Revision 3.4
Page 62
On Delay
0
Off Delay
0
Deadband
1%
Alarm 3
Enabled?
No
Setpoint
60% of scale
Latching
No
Trip On
High
On Delay
0
Off Delay
0
Deadband
1%
Fault Alarm
Setpoint
-10% of scale
Data From
Sensor Type
AI 4-20mA
IO Range
0-20mA
Min Raw
~800
Set at factory, do not adjust
Max Raw
~4000
Set at factory, do not adjust
Filter Samples
5
Local Cal?
No
Marker Enabled
No
Marker %
-15.63
Marker As
IN CAL
Sensor Life
No
Temp Comp
-40
Gain / Offset
1.00 / 0.00
-30
Gain / Offset
1.00 / 0.00
-20
Gain / Offset
1.00 / 0.00
-10
Gain / Offset
1.00 / 0.00
0
Gain / Offset
1.00 / 0.00
+10
Gain / Offset
1.00 / 0.00
+20
Gain / Offset
1.00 / 0.00
+30
Gain / Offset
1.00 / 0.00
+40
Gain / Offset
1.00 / 0.00