| Equipment Set Up | User Guide |
11
14. Connect the power supply to the power
port located on the shoulder control panel.
15. Open the clamps on the A&V in-flow tube
and arterial recirculation tube.
16. Let the blood flow towards the arm.
17. Set the pulse rate control to 80 beats per
minute and pulse strength control to maximum
value.
As the fluid is circulated, air bubbles entrapped
will be flushed from the arterial recirculation
tube to the dispensing bag.
If there is no fluid circulation or the pulse is not
working, please check the tubes to ensure that
there are no kinks and both clamps are open.
If everything is correctly setup and still there is no
circulation,then repeat the priming procedure.
The system is considered primed and ready to
use when the veins are free of air bubbles.