PEDIATRIC HAL®
|
User Guide
Working with Pediatric HAL | 51
Removing the Cartridge
•
Before replacing the cartridge, please ensure that the cartridge is empty.
•
To exhaust the contents of the CO
2
cartridge, turn on CO2 in the software until CO
2
is no longer detected.
Activating CO2 Output
1.
Under the Breathing section, select Lung CO2.
2.
Use the gray slide bar, that ranges from 0-10,
or type in the value to adjust the level of
CO
2
exhaled. To deplete the CO
2
cartridge
faster, raise the level of Lung CO2 to 10.
3.
Securely hold the bottom of the CO
2
cartridge protective case with one hand
and open it by twisting the top counter-
clockwise with the other hand.
4.
Continue to twist the protective
case until the top comes off.