HAL® S1030
|
USER GUIDE
34 | Working with HAL
6.
Click "OK" once the values on the oximeter
monitor and the UNI calibration menu
match. The grey bubble will turn green
and proceed to the next value.
7.
Continue this calibration process and
adjust the values for the oximeter monitor
using the arrows until the process is
done. Once done, click "Finish".
O2 Saturation Usage
After calibrating the oximeter monitor,
proceed with using the O2 Saturation
feature in UNI by doing the following:
1.
Under the 'Breathing' section,
locate 'O2 Saturation'.
2.
Click on 'O2 Saturation' and adjust the slider
bar or type in the desired value that you
want to appear on your oximeter monitor.
3.
Allow the oximeter monitor time
to adjust to the new reading.