GTC-542
Instruction Manual
-
It is a mode that resets only calibration data from the currently saved data to the original factory
-
setting.
-
When "RESET" S/W is pressed, it returns to gas concentration display mode.
- Press "FUNC" S/W to enter.
- By pressing "UP" S/W or "DOWN" S/W, Yes/no status can be set. When it is "Yes", Calibration
-
Initialization function can be used. (Default "no)
- When "RESET" S/W is pressed, it returns to gas concentration display mode.
- It is a function that uses "FUNC" S/W, therefore, there are cases where S/W is pressed
-
repeatedly.
- To prevent this, when "FUNC" KEY is used for Yes/no, FND displays "1" -> "11" -> "111" -> "1111"
-
from the left in 0.5 sec interval.
- When "YES" is selected, initialization is complete and it returns to gas concentration display mode.
- When "no" is selected, it returns to gas concentration display mode without initialization.
- Selecting "RESET" S/W returns to gas concentration display mode without FND display.
6.6.9. Calibration Data Initialization Mode
■ By pressing "FUNC" S/W + "DOWN" S/W and turning ON power, it enters the mode.
■ Calibration Initialization only resets calibration value to factory setting.
▼
▼ ▲
6.6.8. Factory Initialization Mode
■ By pressing "FUNC" S/W + "UP" S/W + "DOWN" S/W and turning ON power, it enters the mode.
■ Factory initial is a mode that resets the data to the original factory setting.
■ In factory initialization mode, most settings are factory preset, therefore, this function should not be adjusted.
When it is unavoidable, the user must seek help from GasTron to adjust.
-
It is a mode that resets the current saved data to the original factory setting.
-
When "RESET" S/W is pressed, it returns to gas concentration display mode.
- Press "FUNC" KEY to enter.
- By pressing "▲" KEY or"▼(Test)" KEY, Yes/ No status can be set and when it is in Test, Factory
-
Initialization Function can be used. (Default "no)
- When "RESET" key is pressed, it returns to gas concentration display mode
- It is a function that uses "FUNC" KEY, therefore, there are cases where KEY is pressed
-
repeatedly.
- To prevent this, when "FUNC" KEY is used for Yes/no, FND displays "1" -> "11" -> "111" from the
-
left in 0.5 sec interval.
- When "YES" is selected, initialization is complete and it returns to gas concentration display mode.
- When "no" is selected, it returns to gas concentration display mode without initialization.
- Selecting "RESET" KEY returns to gas concentration display mode without FND display.
▼
▼ ▲
- A message indicating completion of setting is displayed as "End" for 2 sec then it returns to gas
-
concentration display status.
6. Operation Method
6. Operation Method