10/10/2023

26

Using the Trend Graph

Up to 15 Components can 
be plotted on the Trend 
Graph

Click on 

Ch

 column and a 

color will appear for this 
gas on the Trend Graph

The 100% Y-axis = gas 
measurement range.

     Example :

Red gas = Methane

100% = 50 ppm

Black component (always 
1

st

 gas clicked) shows 

time & gas concentration.

Appendix 6

.

Summary of Contents for GT5000

Page 1: ...the GT5000 Calcmet manuals for full user instructions or contact local representative or Gasmet Requires that the Hazmat Application Library located in the folder C Identification Tool _ SNxxxxx has...

Page 2: ...Gasmet USB Stick All manuals software and documentation And HASP Keys Gasmet Sample Probe with 1 Teflon Tubing white recommended for sampling 2 Tygon tubing Clear recommended for background measureme...

Page 3: ...3 BATTERY HATCH POWER CONNECTION Door Closed Battery AC Power inlet 3Hrs Battery Life 4 Hrs Recharge time...

Page 4: ...tton sets mode of communication with Tablet Battery status indicator 1 Remove rubber dust protection from GT5000 front panel 2 Apply Power 3 Attach Sample Connector Tubing Probe to Sample In 4 Turn th...

Page 5: ...Control switch to communication mode Cable Wifi Bluetooth default 2 In Calcmet EASY Choose the GT5000 Mode of Operation 1 2 GT5000 Setting Calcmet EASY Setting no action required unless troubleshootin...

Page 6: ...EASY on Gasmet Tablet 6 1 Turn on Tablet 2 Double Tap Calcmet Icon with Stylus or Finger 3 Wait 10 20 secs for Calcmet EASY screen last application library used will automatically load _ Default is Ha...

Page 7: ...Calcmet EASY Software 7 Five different screens available each screen is separate not as shown above Views Results Spectrum Trend Residual Background Left Right arrows move between screens...

Page 8: ...y to measure If Hardware status is Not OK is displayed Click on Details Contact Gasmet or representative if Status OK cannot be displayed after waiting further warm up time 8 If error occurs after cli...

Page 9: ...analyzer front panel 2 One single measurement where the pump is not activated can be taken by clicking or 3 Start a continuous measurement where pump is automatically started by clicking 4 Continuous...

Page 10: ...commended 2 Flush Time setting Menu Measuring Times When N2 gas flow is 1 3 l min set Flush time 120 secs 3 Click Background The GT5000 will automatically proceed to the perform background Time clock...

Page 11: ...met or your representative if there is a Fail for either above parameters 3 The CO2 band should not reach the x axis refer line 3 above If this fails _ Repeat Background but first check following a Is...

Page 12: ...ent where the pump is not activated will be taken Used in special sampling situations 2 Start a continuous measurement started by clicking the GT5000 sample pump should start 3 Continuous measurement...

Page 13: ...ed What is normal for ambient air 1 Carbon Dioxide 400 ppm indoors can range 500 1000 ppm 2 Methane 2 ppm 3 Water Vapor 0 5 3 v v depending on ambient temperatures 4 Other gases should be 0 ppm ALL CL...

Page 14: ...s Results 2 Calcmet Easy Software is reporting air sample is not normal and one or more toxic gases have been detected Residual Column turns from Green to Yellow or Red for one or more gases AND ALARM...

Page 15: ...apor IDENTIFY function automatically searches Gasmet reference library 400 gases and reports gas or gases it identifies with gas concentrations The NIST EPA library with 5 000 gases can also be search...

Page 16: ...rary to the Hazmat library 2 Move from Calcmet EASY to EXPERT Part 1 To Add Identified gases into the Hazmat Library Calcmet EASY Screen view Calcmet EXPERT Screen view 1 Re Load Hazmat Application i...

Page 17: ...alcmet EASY Locate EASY on Icon tray HAZMAT Application now includes identified gas Example Toluene Part 2 To Add Identified gases into the Hazmat Library Calcmet EXPERT Screen view Calcmet EASY Scree...

Page 18: ...as cell mirrors Step 1 Take GT5000 to a location with Clean Ambient Air Step 2 Measure continuously this clean ambient air for at least 3 5 mins Step 3 Exit it from Calcmet Software Calcmet Software w...

Page 19: ...time of the instrument is typically between 45 and 120 seconds depending on the gas measurement time When you start measurement at a new sample point let the analyzer run for minimum of 1 3 minutes s...

Page 20: ...d and replace if necessary If the sample gas is warmer than the sample cell ensure that moisture in the sample does not condense inside the analyzer when the gas cools down Allow the analyzer to warm...

Page 21: ...s step several times to check if loss of communication is temporary 3 Exit Calcmet software on Tablet and Restart Calcmet Software Calcmet automatically pairs with the GT5000 at start up After Calcmet...

Page 22: ...are should be changed from the default of 20 seconds The 3 common measurement time settings are 5s 20s 60s 60s is recommended when seeking lowest detection levels and leaving the GT5000 stationary for...

Page 23: ...the box as shown Note saving location for the files that are stored These include 1 Text File 2 Spectra File 3 Background File 4 Log File C SN26816 will change according to the serial number of your...

Page 24: ...ll be stored as a text file Location of saved files is C SNxxxxx CalcmetResults Date Date when measurements taken Text file can be imported into EXCEL using import wizard Appendix 5 C SNxxxxx will cha...

Page 25: ...Software Advanced Mode Checking this boxes selects all spectra in the folder Re analyzing the spectra can be used to further verify the results or using the IDENTIFY function on another test sample Ca...

Page 26: ...n be plotted on the Trend Graph Click on Ch column and a color will appear for this gas on the Trend Graph The 100 Y axis gas measurement range Example Red gas Methane 100 50 ppm Black component alway...

Page 27: ...find any Unknown gases and it is to be interpreted as a Clean Sample refer limitations statement on last page Visual inspection of spectra shows peaks not normal in clean ambient air indicating there...

Page 28: ...no gases identified 1 Change Limit HQI value to 60 00 and re run the IDENTIFY function 2 Uncheck do not search in NIST Library and re run the IDENTIFY function 3 Measure another sample or Open previou...

Page 29: ...m the application library that are not required 4 Save New application In Calcmet Expert Options Application Create New Toluene_Application Using the example of Toluene gas from slide 17 AND adding Et...

Page 30: ...component When using the GT5000 Terra to accurately quantify identified gases it is strongly encouraged that additional quality control steps be initiated to verify further support the gas analysis r...

Reviews: