Construction and Dimensions 

120 mm 

Hydrogen-Cartridge 

Electrode Head 
PTFE 

Socket for 
2 mm Banana plug 

Electrode shaft  
PTFE 
Ø 8 mm 
Length 80 mm 

Contact wire 

Electrode holder  PEEK 

Hydrogen  electrode 

Commissioning  

Troubleshooting 

Make sure all cables are connected correctly. 
Check if the connections are corroded or damp. 
Note the lifetime of the hydrogen source! If necessary, replace the 
cartridge. 

This hydrogen electrode is a reference electrode intended only for 
potential measurement during electrochemical measurements. 

Note: 

The hydrogen source of this reference electrode must be 

activated before first use.  

Safety Information 

The electrode is intended only for the purpose above. 
The manufacturer is not liable for damage resulting 
from improper use. 
Observe the following instructions, otherwise the 
electrode may be damaged or measurements resukts 
be falsified.  
When working with chemicals, all relevant safety 
regulations of the manufacturer and the labarotory 
must be observed. 

Further Informations 

More information can be found on our website 

www.referenceelectrode.info

Operating Instructions 

Intended Use 

Disposal 

The electrode must not be disposed of as household waste. Please 
dispose of the electrode in a separate collection for electrical and 
electronic equipment according to local regulations.  
This also applies to the hydrogen cells.  
If you have further questions, please contact us. 

Subject to technical changes. 

Storage 

After the measurements, rinse the electrode thoroughly with water. 
Place the hydrogen electrode in a liquid (such as measuring 
solution, 1 mol / l hydrochloric acid, 1 mol / l sodium hydroxide 
solution, water) to keep it functional. 

Runtime of Hydrogen Cartridge 

Once the end of runtime has been reached, HydroFlex must no 
longer be stored in liquid.  
Replace the cartridge and activate the reference electrode as 
described in comissioning. 

HydroFlex 

Hydrogen Reference Electrode 
 

Item no.: 81000 

G A S K A T E L  

Gesellschaft für Gassysteme durch Katalyse und Elektrochemie 
mbH 

 

Hollaendische Straße 195—Building M11 
34127 Kassel, Germany 
 
Phone +49 561 59190 
E-Mail [email protected]  
 
Webshop 

www.gaskatel.com

 

Internet   

www.referenceelectrode.info

 

Reviews: