STORAGE & CLEANING 
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept 
clean at all times, care should be taken when cleaning the lens using only a  soft cloth or cotton  swab  with water  or  medical 
alcohol, allowing the lens to fully dry before using the thermometer. Do not submerge any part of the thermometer.   
LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic  mess ages as follows:

 

            ‘Hi’  or ’ Lo’  is displayed when the temperature being measured is outside of the 

settings of HAL and LAL. 

   

  ‘Er2’   is  displayed  when  the  thermometer  is  exposed  to  rapid  changes  in  the  ambient  tempe

rature.  ‘ Er3’   is  displayed 

when  the  ambient  temperature  exceeds  0

  (32°F)  or  +50

  (122°F).  The  thermometer  should  be  allowed  plenty  of 

time (minimum 30 minutes) to stabilize to the working/room temperature.   

Error  5~ 9,  for  all  other  error  mess ages  it  is  necess ary  to  reset  the  thermometer.  To  reset  it,  turn 

the  instrument  off, 

remove  the  battery  and  wait  for  a  minimum  of  one  minute,  reinsert  the  battery  and  turn  on.  If  the  error  mess age 

remains, please c ontact the Service Department for further as sis tance

 

BATTERIES 

The thermometer incorporates visual low battery indication as follows:

 

 

 

 

 

 

BATTERY REPLACEMENT 

1. Pull out the rubber gasket (10) 

2. Loosen the screw and pull out the battery c over 

3. Replace and reinstall with new batteries 

4. Place the battery c over back and fasten the screw

 

5. Plug the rubber gasket (10) to avoid failure of water proof 

    When the ‘ Low Battery’  icon indic ates the battery is low, the battery should be replaced immediately

 with AAA, 1.5V batteries . 

Please note: It is important to turn the ins trument off before replacing the battery otherwise the thermometer may malfunction 

    Dispose of used battery promptly and keep away from c hildren. 

 

SPECIFICATION 

Measurement Range 

-60 to 550 

  (-76 to +1022 ºF) 

Operating Range 

0 ~  50

  (32~ 122 ºF) 

Acc uracy 

Measured value  ± 2% or  ± 2

 

Emis sivity 

0.1 ~  1.0 (default value: 0.95) 

Response T ime  

1 sec 

D:S (Distance:Spot) 

12:1 

Battery Life 

min 14 hours c ontinuous use 

Dimensions 

143.9 x 116.73 x 42.78mm

 

Weight 

180 grams(6.35 oz) including batteries (AAA*2pcs)

 

Note:  Under  the  electromagnetic  field  of  3V/m  from  200  to  700  MHz,  the  maximum  error  is  5

  (9ºF).  The  thermometer  will 

automatically shut off if left idle for more than 15sec. 

   

 

EMC/RFI

: Readings may be affected if the unit is operated within radio frequency electromagnetic  field strength of 

approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.   

 

 

 

 

 

 

‘Battery Exhaus ted’ : 

measurements are not poss ible 

‘Battery  Low’ :  battery  needs  to  be replaced, 

measurements are still pos sible 

‘Battery  OK’ :  meas urements  

are poss ible 

   