BreathSpec®
– Quickstart Manual
43/67
INFORMATION!
To ensure correct measurements it is absolutely necessary to
connect the supplied exhaust tubes
(Gas out and Sample gas
out).
The exhaust tubes (Gas out and Sample gas out)
must be led
separately into the exhaust system and must not be
connected.
The exhaust system must
not generate any negative pressure
.
6.2 Installing the device
1
Unpack the device.
Remove the foam
spacers.
Remove the accessories
boxes.
Lift the device from the
transport box.
WARNING!
Do not deploy the device in areas exposed to explosive substances or
mixtures!
2
Place it on a robust table
top of minimum footprint
size of 60 cm x 70 cm.
Ensure that there is
enough working space at
the rear side of the
device.