GASCOSEEKER 2-500 USER HANDBOOK
IS THE INSTRUMENT TOP PLATE CLEAN AROUND THE
INFRARED LINK?
The instrument data is transmitted to the DEPOT
CHECKER via the infrared link on the top plate. If this is
dirty or obscured then the instrument may not be able to
communicate with the checker.
ACTION:
CLEAN AREA AROUND INFRARED LINK
WITH DRY CLOTH ENSURING ANY
DEBRIS IS REMOVED FROM OPTICAL
RECESS ON LINK.
DO NOT WIPE MATERIAL INTO
RECESS, RATHER BLOW IT OUT.
HAVE YOU FOLLOWED THE DEPOT CHECKER
INSTRUCTIONS?
Instrument will only operate on DEPOT CHECKER if
instructions are followed, particularly with regard to
instrument being OFF when put on checker and MEASURE
button only operated when DEPOT CHECKER requests
it.
ACTION:
SWITCH INSTRUMENT OFF.
READ INSTRUCTIONS FOR CHECKER.
RE-TEST INSTRUMENT FOLLOWING
INSTRUCTIONS EXACTLY.