GASCOSEEKER 2-500 USER HANDBOOK
68
INSTRUMENT DOES NOT OPERATE ON DEPOT
CHECKER
HAS INSTRUMENT BEEN ISSUED TO DEPOT?
Instrument and Depot Checker require to be set up by
technical staff to ensure that they are operating correctly.
ACTION:
CHECK WITH SUPERVISOR THAT THE
INSTRUMENT HAS BEEN ISSUED FOR
USE ON DEPOT CHECKER.
DO OTHER INSTRUMENTS OPERATE ON DEPOT
CHECKER?
There may be a fault on the DEPOT CHECKER. If other
instruments operate correctly on checker then DEPOT
CHECKER is probably correct.
ACTION:
TRY OTHER INSTRUMENT ON
CHECKER. IF SATISFACTORY, WAIT 5
MINUTES AND TRY AGAIN.
REPORT INCIDENT TO SUPERVISOR
EVEN IF INSTRUMENT OPERATES
SECOND TIME.
DOES THE CHECKER HAVE THE CORRECT GASES
AND DO THE CYLINDERS CONTAIN SUFFICIENT
PRESSURE?
The Depot Checker will only operate satisfactorily if it has
the correct gases connected to it and the gas pressures
are correctly set.
ACTION:
CHECK WITH A SUPERVISOR OR
AUTHORISED TESTER
TO ENSURE DEPOT CHECKER IS
OPERATING CORRECTLY.