Section 2: INSTALLATION

 
2.10  Introduction
 

This manual has been compiled not only for the care and maintenance of the Welch Dry 
Vacuum System now in your possession, but as a helpful reference and guide to prevent many 
problems which can occur if used improperly.

2.11  Unpacking
 

Carefully remove the Dry Vacuum System from the shipping carton. Preserve all paperwork for 
future reference. If damage has occurred from shipment a claim must be fi led with the carrier 
immediately; preserve the shipping carton for inspection by the carrier. If you are required to 
communicate with your dealer or Welch Vacuum be sure to include your order numbers for 
quick identifi cation. Do not return the system to the factory without fi rst calling for a returned 
goods number. Call (847) 676-8800.

2.12  System Mounting
 

Rubber feet are attached to the pump. Rubber feet are excellent for applications involving a 
semi-fl exible surface such as a bench top; they help to isolate noise and eliminate creeping. 

2.13  System Location and Environmental Condition
 

The Dry Vacuum System should be located in a clean, dry and well ventilated area. Do not to 
block the ventilation ports located on the metal shell. The pump should be placed where the 
surrounding temperature remains between 10°C and 40°C (50°F and 104°F). Always check 
to insure the location chosen is protected from direct or indirect moisture contact. The pump 
should be located as closely to its system as possible in order to utilize it most effi

  ciently. The 

pump is rated for indoor use only. Maximum altitude 2000 meters. Operating temperature 
range 5°C to 40°C. Maximum relative humidity of 80% for temp up to 31°C decreasing to 50% 
at 40°C. Rated for ±10% of supply voltage. Pollution Degree2, Installation Category II.

   

WARNING

Don’t operate this pump in an atmosphere containing fl ammable or 

explosive gases or vapors

 

 

   

WARNING

The motor is thermally protected and will automatically restart 

unexpectedly when the overload device resets

2.14  Inlet and Outlet Provisions
 

A inlet hose barb which accepts 3/8” I.D. hose makes it easy to connect the system to your 
concentrator, rotary evaporator or gel dryer. Since the vacuum system operates in the viscous 
fl ow regime, the small diameter of the hose will generate minimal conductance loss. For best 
results, Welch recommends the length of the tubing between the pump and the apparatus be 
kept as short as possible. Hose clamps should be used to hold the hose in place. The hose 
barb in the outlet port of the pump accepts 3/8” ID hose. A vent line can be attached to the 
outlet hose barb which allows gases and vapors pumped through the pump to be piped to a 
fume hood. Be sure to call Welch technical service prior to start-up at (847) 676-8800 if you 
have any questions.

   

WARNING

Never Block The Outer Port.  If the outlet is blocked, pressure will 

buildup in the pump which can lead to the pump head bursting creating 

the potential of serious injury. Remove plug from outer port

Summary of Contents for WELCH 2567B-01

Page 1: ...2567B 01 2567B 50 2567C 02 2567C 50 Warning Not recommended for pumping acid base or organic vapors or gases Warning Be sure to properly identify intake and discharge before using pump See section 2...

Page 2: ...ons are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units conform to the SI International system of units of measureme...

Page 3: ...discharge provisions 2 20 Electrical power 2 28 Vacuum connections 2 29 Vacuum gauges 2 30 Traps Section 03 Operation 3 10 Starting procedures 3 13 Leak detection 3 14 Pressure range 3 15 Shutdown pro...

Page 4: ...the potential of the pump head bursting and causing possible injury to personnel in the area 1 17 All electrical products generate heat To avoid serious burns never touch unit during or immediately af...

Page 5: ...e contact The pump should be located as closely to its system as possible in order to utilize it most e ciently The pump is rated for indoor use only Maximum altitude 2000 meters Operating temperature...

Page 6: ...2 23 Identi cation Symbols Power O Power On 2 24 Grounding Instructions This product should be grounded In the event of an electrical short circuit grounding reduces the risk of electric shock by prov...

Page 7: ...only used to indicate vacuum level when operating in the rough vacuum range 0 to 29 9 InHg Dial vacuum gauges give negative pressure that is pressure below atmospheric The reference point for the gaug...

Page 8: ...is if you have droplets coming out of the exhaust port If you see droplets of water Welch recommends you use a cold trap to capture the water before entering the pump or dilute the water vapor stream...

Page 9: ...Leak Detection The importance of eliminating all leaks in a vacuum system is obvious The pump must remove this added volume of leaked gas to maintain the desired vacuum Leaks for these pumps can be l...

Page 10: ...n for the condition of the pump is the determination of its maximum vacuum capability This can be accomplished by blocking of the intake and reading the vacuum level on the gauge See Section 2 8 5 11...

Page 11: ...ll Dimensions L in Inches cm 17 0 43 2 W in Inches cm 7 5 19 1 H in Inches cm 12 0 30 5 Shipping Weight lbs Kg 24 10 9 Shipping Dimensions L in Inches 21 5 W in Inches 12 5 H in Inches 16 Catalog Numb...

Page 12: ...12 6 11 Pumping Speed Curves...

Page 13: ...6 12 Dimensional Drawing 244 0 CORD ASSY 8 25 4X 8 32 UNC 2B THRU 7 09 6 67 9 45 INTAKE NIPPLE FOR 3 8 TUBING MUFFLER 11 67 OUTLET INLET 182 6 3 50 5 38 7 19 296 4 240 0 209 5 9 61 180 0 169 4 88 9 13...

Page 14: ...ice Kit Seal Kit Suction Foot Kit 2567K 03 2567K 04 2500K 04 1 2 2567K 01 Connecting Rod Assembly 2 2 2 2567K 02 Valve Plate Ass y 2 3 2 O Ring Cylinder 2 2 4 2 O Ring Valve Plate 2 2 5 1 1412C Exhaus...

Page 15: ...2567C 50 7 21 System Components available as kit Cat No 2565K 50 vacuum regulator with separator jar Qty 2565K 50 Description for regulator separator jar 1 Jar Assembly 1 Needle Valve Miniature 1 Inst...

Page 16: ...other warranties are expressed or implied The method of executing this warranty servicing adjusting repairing or replacing shall be at the discretion of Welch Ilmvac Vacuum pumps that have been opera...

Reviews: