TR-RC1xx - User Manual
11.2
Restrictions
The following timings and restrictions are applied when commands are
entered. Exceeding these values causes an error to be returned.
The keypad interface prevents invalid values being entered at the
controller.
11.2.1 Continuous mode
The maximum continuous output current is 1.2A for theTR-RC120 and
1.25A for the TR-RC122.
11.2.2 Switched mode
The maximum delay from the trigger input changing state to the output
current switching on or off is 100µs.
The maximum output current is 2A for the TR-RC120 and 10A for the TR-
RC122.
11.2.3
Pulse mode
The maximum pulsed output current is 2A for the TR-RC120 and 10A for
the TR-RC122.
When operating at high currents, even if you set a delay of 0µs, there is a
delay of approximately 4µs.
The following pulse width and duty cycle restrictions apply:
Pulse
output
current
TR-RC120
TR-RC122
≤
1A
100ms max pulse width
100ms max pulse width
> 1A
10ms max pulse width
10ms max pulse width
> 1.5A
50% max duty cycle
50% max duty cycle
> 2A
Not applicable
1ms max pulse width
50% max duty cycle
> 3A
Not applicable
1ms max pulse width
10% max duty cycle
> 5A
Not applicable
1ms max pulse width
7% max duty cycle
> 7A
Not applicable
0.5ms max pulse width
3.5% max duty cycle
—