background image

 

6 | 

P a g e

 

 

3.7. Count function 

 

3.7.1

 

Choose samples. To precisely count the  articles, firstly the quantity of 
samples for counting should be chosen in according with the weights of 
the  articles.  The  quantity  of  samples  can  be  chosen  from 
“1-10-20-50-100”. For a small number of articles, the larger quantity of 
samples, the more precise.   

3.7.2

 

Place the samples on the pan, the balance will show the total weights of 
samples.  Then  press  “Count”,  balance  will  show  “1”,  and  unit  will 
show  “pcs”,  which  indicate  that  balance  is  now  counting.  And  all  the 
samples  are  one  unit.  Then  press  “Count”,  the  displaying  will  show 
between  1-10-20-50-100.  The  quantity  of  articles  must  be  the  same 
with  the  quantity  of  samples,  and  place  the  similar  articles.  The  total 
amount  of  objects  is  showed.  Press  the  “Count”  again  back  to  the 
normal situation. 

 

 

4.

 

Attachment: settings and meanings 

 

Turn  on  the  Scale  by  pressing 

  key,  in  order  to  change  the  configuration  and 

function.  the  display  shows  C1  and  the  relevant  number,  which  can  be  changed. 

Pressing 

  can change the parameters; pressing “TARE” can change the values of 

parameters. 

 

      4.1    C1------ sensitivity setting 0 1 2 3 4, the larger number is, the less sensible but 

the better stable. Default setting is 2 or 1 

4.2

 

  C2------filtering  strength  setting  0  1  2  3,  the  larger  number  is,  the  lower 
reaction speed, but the better stable. Default setting is 2 or 1 

4.3

 

  C3------Baud  rate  setting  2(600)  3(1200)  4(2400)  5(4800)  6(9600).  Default   
setting is 2 

4.4

 

  C4------communication  setting.  “TARE”  changes  the  lower  unit.  “kg/lb” 
changes  the  higher  units.  This  is  the  first  data  during  communication.  Default 
setting is 27. 

4.5

 

Re-pressing 

  after setting. Then the balance is auto calibrated to 0. 

4.6

 

  Press “TARE” and do not loose till “F----3” (Balance should be switched on).   
All the parameters can be reset to default settings. Re-calibration before using. 

 
 

5.

 

   

  Data Transmission 

Balance  is  targeting  data  output  interface  RS232C  and  can  be  directly 
connected to the printer. The following soft wares are needed, when connects to 
computer: 

Summary of Contents for TC50K

Page 1: ...1 41564 Kaarst Germany Operation Manual For Models TC50K TC60K TC100K TC150K TC200K TC300K TC60KA TC75KA TC100KA TC150KA TC300KA To guarantee the best performance of the device please read this operat...

Page 2: ...Specification and technical data 3 Operation method 4 Attachment Configuration and description of function Change 5 Data transmission 6 Structure 7 Caution for Operation 8 Trouble shooting 9 Packing...

Page 3: ...apting to the adverse environment and long hours continuous work With double power supply AC DC With auto calibration With Count function the counting of Object with large quantity simplified Unit con...

Page 4: ...kg 0 150 kg 0 200 kg 0 300 kg Plattform Dimension 520 x 410 mm Volume 970 B x 515 T x 205 H mm Power Supply AC 230V 50Hz 60Hz 4W Rechargeable Battery DC 6V 4Ah 20HR Use Temperature 0 40 Use Humidity 8...

Page 5: ...till press TARE to return to zero 3 3 The units can be changed between kg and lb Press to change the units during weighting 3 4 To use the balance in dark place press the key to active the background...

Page 6: ...to change the configuration and function the display shows C1 and the relevant number which can be changed Pressing can change the parameters pressing TARE can change the values of parameters 4 1 C1 s...

Page 7: ...enter 2unit 7units 3unit 1 unit 1 unit 3 8 4 Check bits no check bits 3 8 5 Output pins content 9 core socket 2 pins RXD 3 pins TXD 5pins GND 3 8 6 Computer can control balance though RS232 The order...

Page 8: ...8 P a g e 6 Structure 1 TARE key 2 Background light key 3 COUNT key 4 Kg lb key for unit change 5 PRINT key 6 ON OFF key 7 Special signal interface 8 RS 232 interface 9 Power supply...

Page 9: ...use of electronic balance without power supply 7 4 Please use separate and individual plug for the balance to avoid any possible interference or disturbance 7 5 Place the scale on a firm table for op...

Page 10: ...e message F H 1 Object weight beyond capacity 3 Error by sensor 1 remove the object 2 restore the fabric configuration The message C L by Calibration Error by sensor restore the fabric configuration T...

Page 11: ...ing the collection systems available If electrical appliances are disposed of in landfills or dumps hazardous substances can leak into the groundwater and get into the food chain damaging your health...

Reviews: