3.

 

Operational Method

 

(1)

 

Turn on the power. Open the switch. The displaying windows displays from 
“F----0”to “F----9” and after a certain time for stabilizing, “

0

” appears; Then 

keep the scale on for 15 minutes. Instability of the displaying at every 
beginning is normal; after a certain time, the displaying will stabilize. 

(2)

 

In case of empty scale platform, the displaying deviates from zero, pressing 
“TARE” key, then the displaying will return to zero. 

(3)

 

If the scale has not been used for a long time or just bought, the scale should 
be calibrated. Firstly, the scale is fully preheated (over 15 min.) with empty 
scale platform. Then press “CAL”, the displaying window displays “CXXX” 
and the balance will automatic calibrate (XXX should be the weight of 
standard weights). At the same time, only place calibrated weights on the 
scale platform. After stabilizing, the scale displays the weight of the 
calibrating weights; till now the calibration is completed and the scale can be 
normally used. If press “CAL”, but display “C----F”, which indicates the Zero 
point is not stable. Then re-press “TARE” to make “zero” point stable; and 
re-press the “CAL” to carry out calibration. 

(4)

 

If weight of a weighed article exceeding the scale measuring range, the scale 
will warn “F----H”. 

(5)

 

If needing to tare, first place the container on the platform, after stabilizing, 
press “TARE” ;the scale displays “zero”. Then place the article in container, 
at this time, it shows the net weight of the article. Remove the article and the 
container. The scale shows a negative value of container weight. Still press 
“TARE” to return to “0” 

(6)

 

The units can be changed between “kg”, “lb”. Press “

” to change the units 

during weighing. 

(7)

 

Count 
First choose samples. To precisely count the articles, the quantity of samples 
for counting should be chosen in according with the weights of the articles. 
The quantity of samples can be chosen from “1-10-20-50-100”. For a small 
number of articles, the larger quantity of samples, the more precise. Place the 
samples on the pan, the balance will show the total weights of samples. Then 
press “COUNT”, balance will show “1”, and unit will show “pcs”, which 
indicate that balance is now counting. And all the samples are one unit. Then 
press “COUNT”, the displaying will show between 1-10-20-50-100. The 
quantity of articles and samples must be the same, and place the similar 
articles. The total amount of objects is showed. Press the “COUNT” again 
back to the normal situation. 

(8)

 

Data Output 
Balance is targeting data output interface RS232C and can be directly 
connected to the printer. 6 feet and 8 feet of RS232C can connect to an 
external print button or to computer. The following soft wares are needed, 