3. Operational

Operational

Operational

Operational Method

Method

Method

Method

3.1 Turn on the power. Open the switch. The displaying windows displays from “F----

1”to “F----9” and after a certain time for stabilizing, “

0

·

0

” appears; Then preheats

for 15 minutes. Instability of the displaying at every beginning is normal; after a
certain time, the displaying will stabilize.

3.2 With empty scale platform, the displaying deviates from zero, pressing “TARE” key,

then the displaying will return to zero.

3.3 If the scale has not been used for a long time or just bought, the scale should be

calibrated. Firstly, the scale is fully preheated (over 15 min.) with empty scale
platform. Then press “CAL”, the displaying window shows “CXXX” and the balance
will automatic calibrate (XXX should be the weight of standard weights). At the same
time, only place calibrating weights on the scale platform. After stabilizing, the scale
displays the weight of the calibrating weights and the unit is “g”; till now the
calibration is completed and the scale can be normally used. If press “CAL”, but
display “C----F”, which indicates the Zero point is not stable. Then re-press “TARE”
to make “zero” point stable; and re-press the “CAL” to carry out calibration.
If weight of a weighed article exceeding the scale measuring range, the scale will

warn”F----H”.

3.4 If needing to tare, first place the container on the platform, after stabilizing, press

“TARE” ;the scale displays ”zero”. Then place the article in container, at this time, it
shows the net weight of the article. Remove the article and the container. The scale
shows a negative value of container weight. Still press “TARE” to return to “zero”

3.5 Count

3.5.1 First choose samples. To precisely count the articles, firstly the quantity of

samples for counting should be chosen in according with the weights of the
articles. The quantity of samples can be chosen from “1-10-20-50-100”. For a
small number of articles, the larger quantity of samples, the more precise.

3.5.2 Place the samples on the pan, the balance will show the total weights of samples.

Then press “Count”, balance will show “1”, and unit will show “pcs”, which
indicates that balance is now counting. And all the samples are one unit. Then
press “Count”, the displaying will show between 1-10-20-50-100. The quantity
of articles and samples must be the same, and place the similar articles. The total
amount of objects is showed. Press the “Count” again back to the normal
situation.

3.6 The units can be changed between “g”, “ct”, “ozt”, “oz”, “GN”, “dwt”, “lb”. Press

“Units” to change the units during weighting.

3.7 Data Output

Balance is targeting data output interface RS232C and can be directly connected to the
printer. 6 feet and 8 feet of RS232C can connect to an external print button or to
computer. The following soft wares are needed, when connects to computer:
3.7.1

Transmission format: serial asynchronous

3.7.2

Serial communication protocol: Baud rat:600, data units: 8 units, starting unit:

1 unit, ending unit: 1unit

Summary of Contents for JJ100Y

Page 1: ...000Y JJ3000Y JJ5000Y JJ6000Y Operation Manual G G GmbH Koenigstr 34 41460 Neuss Germany G G GmbH Content 1 Overview 2 Specifications and Performance indexes 3 Operational Method 4 Outer structure figure of electronic scale 5 Cautions for operation 6 Windscreen and calibration weights 7 The batteries power socket and RS232 data interface 8 The Warranty 9 Contact ...

Page 2: ...no other product can compete with the product is already widely used in machinery industry chemical textile as well as food industry where a high standard of measure and analysis procedure is preferred Many companies and enterprises have improved their quality control procedure with lower cost factor and higher working efficiency by adapting our instrument 2 2 2 2 Specifications Specifications Spe...

Page 3: ... stabilizing press TARE the scale displays zero Then place the article in container at this time it shows the net weight of the article Remove the article and the container The scale shows a negative value of container weight Still press TARE to return to zero 3 5 Count 3 5 1 First choose samples To precisely count the articles firstly the quantity of samples for counting should be chosen in accor...

Page 4: ...ngs meanings meanings 1 Press CAL and do not loose till F 3 Balance should be switched on And pressing CAL can change the parameters pressing TARE can change the values of parameters 2 C1 sensitivity setting 0 1 2 3 4 the larger number is the less sensible but the better stable Default setting is 2 or 1 3 C2 filtering strength setting 0 1 2 3 the larger number is the lower reaction speed but the b...

Page 5: ...icles Avoid impact and heavy load on top never exceed its max weighing limit 5 3 Place the scale on a firm table for operation and make sure that the operating environment of the balance is free of great vibration and interference of power supply and of corrosive gases and liquids 5 4 Ensure the warm up time after power on 5 5 Warning if the instrument is used in a manner not specified in this ins...

Page 6: ...ket socket socket socket and and and and RS232 RS232 RS232 RS232 data data data data interface interface interface interface 7 1 Battery Specification 4 pics C type batteries 7 2 Following is the diagram for battery change 7 3 Here follows the diagram of power socket and RS232 data interface ...

Page 7: ...ects and warranty cases we are entitled at our own discretion to either replace the delivery at the prices which were valid at the time of the complaint or to rectify the defect 8 3 We have the right to try to replace the defective product at most twice There will be a slight charge for the repair service of those products by which the warranty period expires Defective goods may only be returned t...

Page 8: ...he groundwater and get into the food chain damaging your health and well being When replacing old appliances with new ones the retailer is legally obligated to take back your old appliance for disposals at least free of charge 9 9 9 9 contact contact contact contact Address G G GmbH Koenigstr 34 41460 Neuss Germany Tel 49 2131 609 7271 Fax 49 2131 609 7273 Email info gandg de Website www gandg de ...

Reviews: