Calibration – Cable Calibration
40
Procedure to calibrate the cable
1)
Connect the
Chassis Ground
on the back of your potentiostat to a known, good earth ground.
2)
Connect the cell cable to the correct color-coded receptacles on the 2 kΩ Calibration Cell.
3)
Place the Calibration Cell inside the Calibration Shield, close the lid, and connect the black floating-
ground lead of your cell cable to the Shield’s grounding post.