background image

AQUA-GLO 

OPERATING 

PROCEDURE 

MANUAL

BULLETIN 159

(5-19)

GAMMON TECHNICAL PRODUCTS, INC.

P.O.BOX 400 - 2300 HWY 34

MANASQUAN, N.J. 08736

PHONE

FAX

WEBSITE

STORE

732-223-4600

732-223-5778

www.gammontech.com

www.gammontechstore.com

GAMMON

AQUA-GLO SERIES V WATER DETECTOR KIT

OPERATING PROCEDURE MANUAL

MODELS GTP-322 AND GTP-323

ASTM METHOD D3240 & D2276

INTRODUCTION

The Aqua-Glo Series V Water Detector Kit is used to measure the amount of undissolved water in kerosene-type jet fuels. 

It can be used on other fuels by following special procedures in the section titled, “Other Fluids.”

A measured sample of fuel is passed through a treated detector pad or “test pad.” Undissolved (free) water in the fuel will 

react with the test pad. When the pad is subsequently illuminated by ultraviolet light, the coating that was contacted by free 

water will fluoresce (glow) a bright yellow with the brightness increasing for increasing amounts of free water in the fuel 

sample. Water content is read out by the Aqua Glo instrument in parts per million by volume. By varying the fuel sample size, 

the standard range of the instrument of 1 to 12 ppm can be increased to as much as 60 ppm of free water.

METHOD DESCRIPTION

The following procedure is primarily applicable for dynamic line sampling; that is, taking the fuel sample while the fuel 

system is flowing and through the test pad without exposing the sample to the atmosphere or to a sample container. The use 

of sample containers such as bottles or cans for the temporary storage or transport of the sample will result in large errors 

due to changes in temperature, adsorption of water on container walls, etc., and is not recommended. The amount of free 

water in a sample is very sensitive to temperature.

For a simplified step-by-step procedure, go to the end of this manual for the Abbreviated Procedure; first time users should 

read the entire manual first.

CALIBRATION

An encapsulated pad labeled “Calibrating Standard” is provided in a separate container in each kit. Note the “set” number 

on the back of this pad. Place the calibrating standard in the hinged flap window with the blank side (the side without the 

set number) facing the instrument. Turn on the ultraviolet lamp and then press and hold the switch button on the photocell 

comparator (“Instrument Pack”). This button is located within a cup on the instrument. Adjust the light modulating lever until 

the meter pointer is steady at the zero “O” in the center of the scale. Always move the light modulating lever in the same 

direction when zeroing the photocell comparator, when calibrating or testing, to eliminate errors caused by clearance in the 

mechanical linkage. This may take 45 seconds. If the reading obtained on the scale located on the back of the Aqua Glo 

instrument does not agree with the “set” number on the calibrating standard, adjust the photocell comparator on the side 

of the instrument pack (at the 45 degree bend). Insert the included small screwdriver, and adjust as necessary. Repeat 

the above procedure until the reading obtained agrees with the calibrating standard “set” number. 

THE INSTRUMENT 

SHOULD BE CALIBRATED PRIOR TO EACH USE AND AFTER EACH HOUR OF USE.

Return the calibrating standard to its container in the kit. Do not use a calibrating standard from another kit. Each calibrating 

standard is matched to a fluorescing standard which is secured under the panel labeled “Fluorescing Standard.” The code 

number on the Fluorescing Standard must be the same as the code number on the Calibrating Standard 

but the “set” 

points need not be the same

. Instruments that have been manufactured since 1998 have a small decal on the stationery 

face of the hinge providing the code number. NOTE: We urge our customers to label older instruments accordingly so that 

operators can easily determine if they are using the matching calibrating standard.

SAMPLE CONNECTION

The sample connection must be planned so that the flow path to the test pad is as short as possible and so that there are 

no traps or pockets that may collect water. 

Reviews: