Specifications
75
8.2.3
Measuring styli
Only wipe off the styli with a dry, lint-free cloth. If necessary, clean the tips with rubbing alcohol.
Never treat the sockets and the axes with oil or any other lubricants.
8.2.4
Face bow
Parts of the face bow must only be disinfected by spray or wipe disinfection. Wash off any residues un-
der running water. Only the glabella support may be sterilized in an autoclave (5 minutes at 134 °C; 20
minutes at 120 °C).
8.2.5
Functional occlusal clutch
Disinfect the functional occlusal clutch in a disinfecting tank or use disinfectant spray. Subsequently,
clean the registration material off the clutch. Sterilization can be carried out in an autoclave (5 minutes
at 134 °C; 20 Minutes at 120 °C).
8.2.6
Occlusal tray clutch
Disinfect the occlusal tray clutch in a disinfecting tank or use disinfectant spray. Subsequently clean
the registration material off the clutch. Sterilization can be carried out in an autoclave (5 minutes at 134
°C; 20 Minutes at 120 °C).
8.2.7
Bite fork
Disinfect the bite fork in a disinfecting tank or use disinfectant spray. Subsequently clean off any regis-
tration material. If a thermoplastic material has been used, it is advisable to place the bite fork in a refri-
gerator to make removal of the registration material easier. Sterilization can be carried out in an auto-
clave (5 minutes at 134 °C; 20 Minutes at 120 °C).
8.2.8
3D joint support
The 3D joint support must only be disinfected by wipe disinfection.
Never treat the joints of the 3D joint support with oil or any other lubricants and do not place the
part in an immersion bath.
8.3
Recurrent tests
Like all other electronic devices, also the CADIAX device and its electronic components go through the
process of aging. Therefore, the measurement function and security of the device has to be tested at
regular intervals. A test interval of one year is recommended.