SUPPORTS - pag. SU 8
Rotating instrument carrying tray
ATTENTION:
Fro a correct and safe use of the rotating tray with MINI it is necessary to read
carefu
lly the text container in sections “INFORMATION” (from pag. IN 1) and
“BODY” (from pag. BO 1) of the usage instructions.
The rotating instrument carrying tray must be placed in the special space on the
body of the trolley (see lay-out on page SU1).
Failure to observe the prescriptions and the procedures described in this
document absolves the manufacturer of any responsibility for damages to things
or people.
Function
This accessory shall be used
as a support for defibrillators or
monitors.
Features
The surface of the tray is equipped with an anti-sliding pad.
Dimensions: 38x31 cm
Usage
ATTENTION
Improper use of the rotating tray may cause damages to the defibrillator.
Make sure you do not exceed the maximum allowable loading weight of 12 kg
(reported in the label that is stuck to the base of the tray surface).
The dimensions of the defibrillator must be compatible with the tray.
Do not lean on or pull the defibrillator to move the trolley.
Do not lay additional weights onto the defibrillator.
When you move the trolley, the tray shall be moved back to the safety position
–
within the trolley profile
1
Lift the specific bushing, Rotate the surface into the desired lockable position.
2
Lay the defibrillator within tray borders.
Removal
Follow the procedure described for assembly in reverse order.
Maintenance
Refer to what is reported on pag. IN 12.
Decommissioning and disposal
Refer to what is reported on pag. IN 14.