NOTE: 

If you want to take another reading, you can rein-

flate up to twice. Wait at least 2- 3 minutes before measur-
ing again, to avoid inaccurate readings due to the engorged 
blood vessels.

QUICK REFERENCE GUIDE

BEFORE

1.  Have the patient to RELAX for at least 5 minutes before 
taking a measurement.
2.  Remove tight fitting clothing from the upper arm.
3.  The patient should NOT eat, smoke or exercise for at 
least 30 minutes before taking a measurement.

DURING

1.  PLACE CUFF at the heart level.
2.  INFLATE unit to proper level. (20-30 mmHg above esti-
mated systolic pressure).
3.  The patient should not talk or move during a measure-
ment .
4.  DEFLATE unit at 2-4 mmHg per second.

AFTER 

1.  WAIT 2-3 minutes before taking another measurement.

CARE AND MAINTENANCE

To protect your unit from damage, please AVOID washing or 
moistening the cuff or dropping the gauge.

Your new blood pressure unit has been carefully checked to 
assure reliability and accuracy prior to shipment and use.
However, as with any sensitive instrument subjected to 
repeated use, we recommend that your blood pressure 
gauge be cheeked periodically. When the cuff is fully 
deflated, the gauge needle must stay within the accuracy 
indicator zone. If the needle points outside of the accuracy 
indicator zone, the gauge will give inaccurate readings. 

In 

this case you have to bring the device to the nearest service centre 

for calibration (address mentioned in warranty card).

IMPORTANT

:CONSULT A DOCTOR TO DETERMINE THE  

ATIENTS NORMAL BLOOD PRESSURE. ONLY A  CERTIFIED 
HEALTHCARE PRACTITIONER IS QUALIFIED TO DIAGNOSE 
AND TREAT HIGH BLOOD PRESSURE. PRACTICE THE 
MEASUREMENT PROCEDURES FOR  THIS INSTRUMENT 
CAREFULLY.