Gamma 700 Instruction Manual Download Page 4

6

7

5. Rest their arm on a table. The arm should be relaxed with 
the palm turned upward. Ensure the cuff is same level as 
heart.

6.  Hold the gauge in one hand with the numbers directly in  
front of you. Hold the inflation bulb in your other hand. Close 
the air release valve attached to the inflation bulb by turning 
it clock wise.

7.  To inflate the cuff, rapidly squeeze the inflation bulb. 
Inflate the cuff until the gauge reaches approximately 20-30 
mmHg ABOVE normal SYSTOLIC reading.

CAUTION: 

DO NOT INFLATE THE CUFF ABOVE 280 mmHg.YOU MAY 
INJURE THE PATIENT OR DAMAGE THE INSTRUMENT.

8. Slowly open the air release  valve by turning it coun-
ter- clockwise. Release the air at a rate of 2-4 mmHg per 
second.

9.   Listen carefully for the appearance of sounds, watching 
the gauge needle. When you hear the first appearance of 
sound, it is the SYSTOLIC blood pressure reading.

10.  Continue to release the air at a rate of 2-4 mmHg per 
second. When you no longer hear any sounds, that is  the 
DIASTOLIC blood pressure reading.

11. Turn the air release valve counter-clockwise to release 
the remaining air. Record the reading.

Reviews: